Your browser doesn't support javascript.
loading
Accuracy of a novel method for IOP measurement without applying pressure on the cornea.
Biran, Amit; Sayar, Mordehai; Lazzar, Moshe; Mimouni, Michael; Beiran, Itzchak.
Afiliação
  • Biran A; Department of Ophthalmology, Soraski Medical Center, Tel Aviv, Israel.
  • Sayar M; Department of Applied Physics, Weizmann Institute of Science, Rehovot, Israel.
  • Lazzar M; Department of Ophthalmology, Soraski Medical Center, Tel Aviv, Israel.
  • Mimouni M; Department of Ophthalmology, Rambam Health Care Campus Affiliated With the Bruce and Ruth Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, POB 9602, 3109601, Haifa, Israel. michael@intername.co.il.
  • Beiran I; Department of Ophthalmology, Rambam Health Care Campus Affiliated With the Bruce and Ruth Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, POB 9602, 3109601, Haifa, Israel.
Int Ophthalmol ; 44(1): 300, 2024 Jul 01.
Article em En | MEDLINE | ID: mdl-38951418
ABSTRACT

PURPOSE:

To assess the accuracy of a prototype novel instrument for intra ocular pressure (IOP) measurements not involving corneal pressure application.

DESIGN:

Prospective case control study.

METHODS:

An institutional study including 16 healthy volunteers without ocular pathology. IOP in both eyes of the participants was measured four times in different body positions with the novel prototype and reference instrument (Goldmann applanation tonometer (GAT) or iCare (iCare Finland OY, Vantaa, Finland)). IOP results were compared between the prototype and the reference instruments in 116 pairs of measurement.

RESULTS:

Overall no statistically significant difference was found between the presented prototype and the reference instrument. Stratifying measurements by instrument used revealed no significant difference for GAT and statistical significant (yet clinically insignificant) difference for iCare.

CONCLUSIONS:

The presented prototype demonstrates good clinical agreement of IOP measuring results with reference instruments Further large-scale studies assessing this instrument in glaucoma patients are warranted.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Córnea / Pressão Intraocular Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Córnea / Pressão Intraocular Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Int Ophthalmol Ano de publicação: 2024 Tipo de documento: Article