Your browser doesn't support javascript.
loading
Hybrid convergent procedure with proactive oesophageal cooling for the treatment of long-standing persistent atrial fibrillation: a case series.
Velasco, Alejandro; Buch, Chirag; Hui, Dawn; Joseph, Christopher; Onsager, David; Zagrodzky, William; Kulstad, Erik; Nayak, Hemal M.
Afiliação
  • Velasco A; Division of Cardiology, University of Texas Health, San Antonio, TX, USA.
  • Buch C; Division of Cardiology, University of Texas Health, San Antonio, TX, USA.
  • Hui D; Division of Cardiothoracic Surgery, University of Texas Health, San Antonio, TX, USA.
  • Joseph C; Department of Emergency Medicine, University of Texas Southwestern Medical Centre, Dallas, TX, USA.
  • Onsager D; Division of Cardiothoracic Surgery, University of Chicago Medicine, Chicago, IL, USA.
  • Zagrodzky W; Department of Biochemistry, Colorado College, 14 E Cache La Poudre St, Colorado Springs, CO 80903, USA.
  • Kulstad E; Department of Emergency Medicine, University of Texas Southwestern Medical Centre, Dallas, TX, USA.
  • Nayak HM; Division of Cardiology, University of Texas Health, San Antonio, TX, USA.
Eur Heart J Case Rep ; 8(7): ytae301, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38966596
ABSTRACT

Background:

The hybrid convergent procedure is approved to treat symptomatic patients with long-standing persistent atrial fibrillation (AF). Despite direct visualization during surgical ablation as well as the use of luminal oesophageal temperature (LET) monitoring, oesophageal injury is still possible. A dedicated device for proactive oesophageal cooling has recently been cleared by the Food and Drug Administration to reduce the likelihood of ablation-related oesophageal injury resulting from radiofrequency cardiac ablation procedures. This report describes the first uses of proactive oesophageal cooling for oesophageal protection during the epicardial ablation portion of hybrid convergent procedures. Case

summary:

Five patients with long-standing persistent AF underwent hybrid convergent ablations with the use of proactive oesophageal cooling as means of oesophageal protection. All cases were completed successfully with no adverse effects. Most notably, cases were shorter when compared to cases using LET monitoring, likely due to lack of pauses for overheating of the oesophagus that would otherwise be required to prevent damage to the oesophagus.

Discussion:

This report describes the first uses of proactive oesophageal cooling for oesophageal protection during the epicardial ablation portion of five hybrid convergent procedures. Use of cooling enabled uninhibited deployment of lesions without the need to pause energy delivery due to elevated temperatures in the oesophagus, providing a feasible alternative to LET monitoring.