Your browser doesn't support javascript.
loading
Development of an Integrated Disposable Device for SARS-CoV-2 Nucleic Acid Extraction and Detection.
Ma, Jing; Hao, Yan Zhe; Hou, Mei Ling; Zhang, Xiao Shan; Liu, Jing Duan; Meng, Hao di; Chang, Jiang Bo; Ma, Xue Jun; Liu, Ji Hua; Ying, Qing Jie; Wang, Xian Hua; Li, Hong Xia; Cao, Yu Xi; Zhang, Xiao Guang.
Afiliação
  • Ma J; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Hao YZ; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Hou ML; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Zhang XS; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Liu JD; Beijing Runbio Biotechnology Development Co., Ltd, Beijing 100106, China.
  • Meng HD; Beijing Runbio Biotechnology Development Co., Ltd, Beijing 100106, China.
  • Chang JB; Beijing Beta Technology Co., Ltd, Beijing 101318, China.
  • Ma XJ; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Liu JH; Beijing Runbio Biotechnology Development Co., Ltd, Beijing 100106, China.
  • Ying QJ; Jiangsu Qitian Gene Biotechnology Co., Ltd, Wuxi 214135, Jiangsu, China.
  • Wang XH; Beijing Beta Technology Co., Ltd, Beijing 101318, China.
  • Li HX; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Cao YX; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
  • Zhang XG; National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China.
Biomed Environ Sci ; 37(6): 639-646, 2024 Jun 20.
Article em En | MEDLINE | ID: mdl-38988114
ABSTRACT

Objective:

To develop a highly sensitive and rapid nucleic acid detection method for the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).

Methods:

We designed, developed, and manufactured an integrated disposable device for SARS-CoV-2 nucleic acid extraction and detection. The precision of the liquid transfer and temperature control was tested. A comparison between our device and a commercial kit for SARS-Cov-2 nucleic acid extraction was performed using real-time fluorescence reverse transcription polymerase chain reaction (RT-PCR). The entire process, from SARS-CoV-2 nucleic acid extraction to amplification, was evaluated.

Results:

The precision of the syringe transfer volume was 19.2 ± 1.9 µL (set value was 20), 32.2 ± 1.6 (set value was 30), and 57.2 ± 3.5 (set value was 60). Temperature control in the amplification tube was measured at 60.0 ± 0.0 °C (set value was 60) and 95.1 ± 0.2 °C (set value was 95) respectively. SARS-Cov-2 nucleic acid extraction yield through the device was 7.10 × 10 6 copies/mL, while a commercial kit yielded 2.98 × 10 6 copies/mL. The mean time to complete the entire assay, from SARS-CoV-2 nucleic acid extraction to amplification detection, was 36 min and 45 s. The detection limit for SARS-CoV-2 nucleic acid was 250 copies/mL.

Conclusion:

The integrated disposable devices may be used for SARS-CoV-2 Point-of-Care test (POCT).
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: RNA Viral / Equipamentos Descartáveis / SARS-CoV-2 / COVID-19 Limite: Humans Idioma: En Revista: Biomed Environ Sci Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: RNA Viral / Equipamentos Descartáveis / SARS-CoV-2 / COVID-19 Limite: Humans Idioma: En Revista: Biomed Environ Sci Ano de publicação: 2024 Tipo de documento: Article