Stationary prospective cardiac gated computed tomography-dynamic study in phantoms andin vivo.
Billingsley, Alex; Inscoe, Christina; Attia, Mohamed Fathy; Lu, Jianping; Zhou, Otto; Lee, Yueh Z.
Afiliação
  • Billingsley A; Joint Department of Biomedical Engineering, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States of America.
  • Inscoe C; North Carolina State University, Raleigh, NC 27695, United States of America.
  • Attia MF; Department of Physics & Astronomy, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States of America.
  • Lu J; Eshelman School of Pharmacy, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States of America.
  • Zhou O; Department of Physics & Astronomy, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States of America.
  • Lee YZ; Department of Physics & Astronomy, The University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, United States of America.
Phys Med Biol ; 69(16)2024 Aug 08.
Article em En | MEDLINE | ID: mdl-38996425
ABSTRACT
Objective.This study explores the feasibility of a stationary gantry cardiac gated computed tomography (CT) with carbon nanotube (CNT) linear x-ray source arrays.Approach.We developed a stationary gantry CT system utilizing multipixel CNT x-ray sources. Given the advantages of straightforward x-ray pulse control with these sources, we investigated the potential for gated prospective imaging. We implemented prospective respiratory and cardiac gating control and evaluated the system through dynamic phantom imaging studies followed by imaging of a porcine model.Main Results.The findings revealed minimal anatomical motion artifacts in the heart and lungs, confirming successful physiologic gated acquisition in stationary gantry cardiac CT. This indicates the potential of this imaging approach for reducing artifacts and improving image quality.Significance.This study demonstrates the feasibility of prospective physiological gating with CNT x-ray sources in a stationary gantry setup for cardiac imaging. This approach could potentially alleviate the need for beta blocker administration during cardiac CT scans, thereby increasing the flexibility of the imaging system and enabling the imaging of a wider variety of patient cardiac conditions.