Improved abdominal T1 weighted imaging at 0.55T.
Tasdelen, Bilal; Lee, Nam G; Cui, Sophia X; Nayak, Krishna S.
Afiliação
  • Tasdelen B; Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
  • Lee NG; Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
  • Cui SX; Siemens Medical Solutions USA, Los Angeles, California, USA.
  • Nayak KS; Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
Magn Reson Med ; 92(6): 2580-2587, 2024 Dec.
Article em En | MEDLINE | ID: mdl-38997798
ABSTRACT

PURPOSE:

Breath-held fat-suppressed volumetric T1-weighted MRI is an important and widely-used technique for evaluating the abdomen. Both fat-saturation and Dixon-based fat-suppression methods are used at conventional field strengths; however, both have challenges at lower field strengths (<1.5T) due to insufficient fat suppression and/or inadequate resolution. Specifically, at lower field strengths, fat saturation often fails due to the short T1 of lipid; and Cartesian Dixon imaging provides poor spatial resolution due to the need for a long ΔTE, due to the smaller Δf between water and lipid. The purpose of this work is to demonstrate a new approach capable of simultaneously achieving excellent fat suppression and high spatial resolution on a 0.55T whole-body system.

METHODS:

We applied 3D stack-of-spirals Dixon imaging at 0.55T, with compensation of concomitant field phase during reconstruction. The spiral readouts make efficient use of the requisite ΔTE. We compared this with 3D Cartesian Dixon imaging. Experiments were performed in 2 healthy and 10 elevated liver fat volunteers.

RESULTS:

Stack-of-spirals Dixon imaging at 0.55T makes excellent use of the required ΔTE, provided high SNR efficiency and finer spatial resolution (1.7 × 1.7 × 5 mm3) compared Cartesian Dixon (3.5 × 3.5 × 5 mm3), within a 17-s breath-hold. We observed successful fat suppression, and improved definition of structures such as the liver, kidneys, and bowel.

CONCLUSION:

We demonstrate that high-resolution single breath-hold volumetric abdominal T1-weighted imaging is feasible at 0.55T using spiral sampling and concomitant field correction. This is an attractive alternative to existing Cartesian-based methods, as it simultaneously provides high-resolution and excellent fat-suppression.