Pilot Assessment of Piezoelectric Transducers as a Cost-Effective Alternative to EndoPAT.
J Vasc Res
; 61(5): 252-259, 2024.
Article
em En
| MEDLINE
| ID: mdl-39074454
ABSTRACT
INTRODUCTION:
Current microvascular assessments may not be practical or accessible requiring experienced personnel and/or ongoing equipment costs. Piezoelectric transducers can reliably obtain finger blood pressure waves, similar to peripheral arterial tonometry devices; thus, they could be used to estimate microvascular function. We aimed to validate piezoelectric transducers as an alternative measure of microvascular function compared to EndoPAT.METHODS:
Twenty-five adults (aged 20-64 years) completed reactive hyperemia (5 min forearm circulatory occlusion and 3 min recovery) with piezoelectric transducers on the middle fingers and EndoPAT probes on the index fingers. Average area under the curve (AUC) of the pulse wave signal for the occluded and control arms was determined at baseline, every 30 s post-occlusion, and 10 s around the peak response. Microvascular function index (MFI) was calculated as the ratio of AUC post-occlusion to AUC baseline in the test arm, then normalized to the same ratio in the control arm. MFI at each time point was correlated with the reactive hyperemia index (RHI) from the EndoPAT.RESULTS:
The greatest significance was found between RHI and MFI at 10 s around the peak response (Spearman's r = 0.67, p = 0.0002; Pearson's r = 0.76, p = 0.00001).CONCLUSION:
MFI is a reusable and user-friendly microvascular function assessment that could provide better access to vascular health screening.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Fluxo Sanguíneo Regional
/
Valor Preditivo dos Testes
/
Análise Custo-Benefício
/
Desenho de Equipamento
/
Dedos
/
Hiperemia
/
Microcirculação
Limite:
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
J Vasc Res
Ano de publicação:
2024
Tipo de documento:
Article