Paradoxical generator impedance behavior during catheter ablation in a patient with severe polycythemia.
Sekihara, Takayuki; Nakano, Tomoaki; Yoshida, Akira; Oka, Takafumi; Sakata, Yasushi.
Afiliação
  • Sekihara T; Department of Cardiology, Faculty of Medical Sciences University of Osaka Osaka Japan.
  • Nakano T; Department of Cardiology, Faculty of Medical Sciences University of Osaka Osaka Japan.
  • Yoshida A; Department of Cardiology, Faculty of Medical Sciences University of Osaka Osaka Japan.
  • Oka T; Department of Cardiology, Faculty of Medical Sciences University of Osaka Osaka Japan.
  • Sakata Y; Department of Cardiology, Faculty of Medical Sciences University of Osaka Osaka Japan.
J Arrhythm ; 40(4): 1045-1048, 2024 Aug.
Article em En | MEDLINE | ID: mdl-39139905
ABSTRACT
An extremely high generator impedance in the blood pool can be observed in a patient with severe polycythemia. However, ablation can be performed safely as long as the generator impedance during contact with the myocardial tissue is within acceptable limits.