Correlation between local instantaneous dose rate and oxygen pressure reduction during proton pencil beam scanning irradiation.
Kanouta, Eleni; Johansen, Jacob Graversen; Poulsen, Sara; Kristensen, Line; Sørensen, Brita Singers; Grau, Cai; Busk, Morten; Poulsen, Per Rugaard.
Afiliação
  • Kanouta E; Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.
  • Johansen JG; Department of Clinical Medicine, Faculty of Health, Aarhus University, Aarhus, Denmark.
  • Poulsen S; Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.
  • Kristensen L; Department of Clinical Medicine, Faculty of Health, Aarhus University, Aarhus, Denmark.
  • Sørensen BS; Department of Physics and Astronomy, Aarhus University, Aarhus, Denmark.
  • Grau C; Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.
  • Busk M; Department of Clinical Medicine, Faculty of Health, Aarhus University, Aarhus, Denmark.
  • Poulsen PR; Department of Experimental Clinical Oncology, Aarhus University Hospital, Aarhus, Denmark.
Phys Imaging Radiat Oncol ; 31: 100614, 2024 Jul.
Article em En | MEDLINE | ID: mdl-39157294
ABSTRACT
Background and

purpose:

Oxygen dynamics may be important for the tissue-sparing effect observed at ultra-high dose rates (FLASH sparing effect). This study investigated the correlation between local instantaneous dose rate and radiation-induced oxygen pressure reduction during proton pencil beam scanning (PBS) irradiations of a sample and quantified the oxygen consumption g-value. Materials and

methods:

A 0.2 ml phosphorescent sample (1 µM PtG4 Oxyphor probe in saline) was irradiated with a 244 MeV proton PBS beam. Four irradiations were performed with variations of a PBS spot pattern with 5 × 7 spots. During irradiation, the partial oxygen pressure (pO2) was measured with 4.5 Hz temporal resolution with a phosphorometer (Oxyled) that optically excited the probe and recorded the subsequently emitted light. A calibration was performed to calculate the pO2 level from the measured phosphorescence lifetime. A fiber-coupled scintillator simultaneously measured the instantaneous dose rate in the sample with 50 kHz sampling rate. The oxygen consumption g-value was determined on a spot-by-spot level and using the total pO2 change for full spot pattern irradiation.

Results:

A high correlation was found between the local instantaneous dose rate and pO2 reduction rate, with a correlation coefficient of 0.96-0.99. The g-vales were 0.18 ± 0.01 mmHg/Gy on a spot-by-spot level and 0.17 ± 0.01 mmHg/Gy for full spot pattern irradiation.

Conclusions:

The pO2 reduction rate was directly related to the local instantaneous dose rate per delivered spot in PBS deliveries. The methodology presented here can be applied to irradiation at ultra-high dose rates with modifications in the experimental setup.