Treatment Reproducibility in Brain Stereotactic Radiotherapy Using a Shim Mask Versus a Standard Mask.
Cureus
; 16(8): e66108, 2024 Aug.
Article
em En
| MEDLINE
| ID: mdl-39229440
ABSTRACT
Introduction This study aimed to evaluate the setup accuracy of the new shim mask with mouth bite compared to the standard full brain mask in stereotactic radiosurgery (SRS) and radiotherapy (SRT) treatments for brain metastases or tumors. Method A combined retrospective and prospective design was employed, involving 40 patients treated at our center. Patients previously treated using standard head masks formed the retrospective cohort, while those treated with the Shim mask and mouth bite formed the prospective cohort. Daily cone-beam computed tomography (CBCT) scans were obtained before each treatment session to ensure patient setup accuracy. Key metrics included absolute shifts in translational and rotational directions, the number of repeat CBCTs, and the time interval between CBCTs. Results The Shim mask significantly reduced the mean setup errors in the lateral translation (p=0.022) from 0.17 cm (SD=0.10) to 0.10 cm (SD=0.10), and in X-axis rotation (p=0.030) from 0.79° (SD=0.43) to 0.47° (SD=0.47). By considering cutoff points of 1 mm in translational and 1° in rotational directions, the Shim mask was significantly more accurate in the lateral direction (p=0.004). Moreover, while 70% of patients in the standard group required repeat CBCT scans, none in the Shim group did, resulting in an average time saving of 10.4 minutes per patient. Conclusion The Shim mask with mouth bite offers enhanced immobilization accuracy in SRT/SRS treatments, leading to time and potential cost savings by reducing the need for repeat CBCT scans. This underscores the importance of adopting innovative immobilization techniques to optimize patient outcomes.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Cureus
Ano de publicação:
2024
Tipo de documento:
Article