Development and assessment of case-specific physical and augmented reality simulators for intracranial aneurysm clipping.
Civilla, Lorenzo; Dodier, Philippe; Palumbo, Maria Chiara; Redaelli, Alberto C L; Koenigshofer, Markus; Unger, Ewald; Meling, Torstein R; Velinov, Nikolay; Rössler, Karl; Moscato, Francesco.
Afiliação
  • Civilla L; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Waehringer Guertel 18-20, Vienna, 1090, Austria.
  • Dodier P; Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
  • Palumbo MC; Ludwig Boltzmann Institute for Cardiovascular Research, Vienna, Austria.
  • Redaelli ACL; Department of Neurosurgery, Medical University Vienna, Vienna, Austria.
  • Koenigshofer M; Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
  • Unger E; Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
  • Meling TR; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Waehringer Guertel 18-20, Vienna, 1090, Austria.
  • Velinov N; Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Waehringer Guertel 18-20, Vienna, 1090, Austria.
  • Rössler K; Department of Neurosurgery, Rigshospitalet University Hospital, Copenhagen, Denmark.
  • Moscato F; Clinics of Neurosurgery, University Hospital Pirogov, Medical University of Sofia, Sofia, Bulgaria.
3D Print Med ; 10(1): 30, 2024 Sep 18.
Article em En | MEDLINE | ID: mdl-39292343
ABSTRACT

BACKGROUND:

Microsurgical clipping is a delicate neurosurgical procedure used to treat complex Unruptured Intracranial Aneurysms (UIAs) whose outcome is dependent on surgeon's experience. Simulations are emerging as excellent complements to standard training, but their adoption is limited by the realism they provide. The aim of this study was to develop and validate a microsurgical clipping simulator platform.

METHODS:

Physical and holographic simulators of UIA clipping have been developed. The physical phantom consisted of a 3D printed hard skull and five (n = 5) rapidly interchangeable, perfused and fluorescence compatible 3D printed aneurysm silicone phantoms. The holographic clipping simulation included a real-time finite-element-model of the aneurysm sac, allowing interaction with a virtual clip and its occlusion. Validity, usability, usefulness and applications of the simulators have been assessed through clinical scores for aneurysm occlusion and a questionnaire study involving 14 neurosurgical residents (R) and specialists (S) for both the physical (p) and holographic (h) simulators by scores going from 1 (very poor) to 5 (excellent).

RESULTS:

The physical simulator allowed to replicate successfully and accurately the patient-specific anatomy. UIA phantoms were manufactured with an average dimensional deviation from design of 0.096 mm and a dome thickness of 0.41 ± 0.11 mm. The holographic simulation executed at 25-50 fps allowing to gain unique insights on the anatomy and testing of the application of several clips without manufacturing costs. Aneurysm closure in the physical model evaluated by fluorescence simulation and post-operative CT revealed Raymond 1 (full) occlusion respectively in 68.89% and 73.33% of the cases. For both the simulators content validity, construct validity, usability and usefulness have been observed, with the highest scores observed in clip selection usefulness Rp=4.78, Sp=5.00 and Rh=4.00, Sh=5.00 for the printed and holographic simulators.

CONCLUSIONS:

Both the physical and the holographic simulators were validated and resulted usable and useful in selecting valid clips and discarding unsuitable ones. Thus, they represent ideal platforms for realistic patient-specific simulation-based training of neurosurgical residents and hold the potential for further applications in preoperative planning.