Errors arising from incorrect orientation of E test strips.
J Clin Microbiol
; 33(7): 1966-7, 1995 Jul.
Article
em En
| MEDLINE
| ID: mdl-7665686
Major errors arise if E test strips are placed upside down. Asymmetric zones, or no zone at all, may result. MICs indicated by upside-down tests were almost always considerably higher than true values. This situation differs markedly from that for conventional testing, where orientation of disks is not important.