Your browser doesn't support javascript.
loading
Errors arising from incorrect orientation of E test strips.
Hamilton-Miller, J M; Shah, S; Yam, T S.
Afiliação
  • Hamilton-Miller JM; Department of Medical Microbiology, Royal Free Hospital School of Medicine, London, United Kingdom.
J Clin Microbiol ; 33(7): 1966-7, 1995 Jul.
Article em En | MEDLINE | ID: mdl-7665686
Major errors arise if E test strips are placed upside down. Asymmetric zones, or no zone at all, may result. MICs indicated by upside-down tests were almost always considerably higher than true values. This situation differs markedly from that for conventional testing, where orientation of disks is not important.