Your browser doesn't support javascript.
loading
Technical note: magnetic resonance imaging of agarose gel phantom for assessment of three-dimensional dose distribution in linac radiosurgery.
Rousseau, J; Gibon, D; Sarrazin, T; Doukhan, N; Marchandise, X.
Afiliação
  • Rousseau J; Laboratoire de Biophysique, Faculté de Médecine, Lille, France.
Br J Radiol ; 67(799): 646-8, 1994 Jul.
Article em En | MEDLINE | ID: mdl-8061999
ABSTRACT
An agarose gel phantom is used to evaluate the spatial distribution of the dose delivered by a linac radiosurgery device. Dependence of the absorbed dose on the T1 relaxation time is checked. T1 magnetic resonance images show the close correspondence between the actual absorbed dose distribution and the dose distribution expected by the treatment planning.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Radiocirurgia Idioma: En Revista: Br J Radiol Ano de publicação: 1994 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Radiocirurgia Idioma: En Revista: Br J Radiol Ano de publicação: 1994 Tipo de documento: Article