Performance characteristics of the Pantak DXT-300 kilovoltage X-ray treatment machine.
Aukett, R J; Thomas, D W; Seaby, A W; Gittins, J T.
Afiliação
  • Aukett RJ; Leicestershire Medical Physics Department, Leicester Royal Infirmary NHS Trust, UK.
Br J Radiol ; 69(824): 726-34, 1996 Aug.
Article em En | MEDLINE | ID: mdl-8949675
ABSTRACT
The beam and performance characteristics of a new orthovoltage X-ray unit, the Pantak DXT-300 have been evaluated. Data were collated for four qualities 3.27 mmAl, 7.15 mmAl, 1.65 mmCu and 3.51 mmCu half value layer (HVL) (SE = 0.04 mm). Parameters which were investigated included beam quality, central axis depth dose, relative output, backscatter factors, field uniformity, peripheral dose and head leakage. The calibration procedure and the performance of the dosimetry system have also been described.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Terapia por Raios X Limite: Humans Idioma: En Revista: Br J Radiol Ano de publicação: 1996 Tipo de documento: Article