High-performance liquid chromatographic analysis of methadone hydrochloride oral solution.
J Pharm Sci ; 66(12): 1749-51, 1977 Dec.
Article em En | MEDLINE | ID: mdl-925941
ABSTRACT
A direct and rapid high-performance liquid chromatographic assay for methadone hydrochloride in a flavored oral solution dosage form is described. A syrup sample, one part diluted with three parts of water, is introduced onto a column packed with octadecylsilane bonded on 10 micrometer porous silica gel (reversed phase). A formic acid-ammonium formate-buffered mobile phase is linear programmed with acetonitrile. The absorbance is monitored continuously at 280 or 254 nm, using a flow-through, UV, double-beam photometer. An aqueous methadone hydrochloride solution is used for external standardization. The relative standard deviation was not more than 1.0%. Drug recovery from a syrup base was better than 99.8%.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Metadona Idioma: En Revista: J Pharm Sci Ano de publicação: 1977 Tipo de documento: Article
Buscar no Google