Biocompatibility of a new device of self-expandable covered and non-covered tracheal stent: comparative study in rats
Rodrigues, Olavo Ribeiro; Minamoto, Hélio; Canzian, Mauro; Correia, Aristides Tadeu; Jatene, Fabio Biscegli.
Afiliação
  • Rodrigues, Olavo Ribeiro; Universidade de Mogi das Cruzes. Divisão de Cirurgia Torácica Geral. Mogi das Cruzes. Brasil
  • Minamoto, Hélio; Universidade de Mogi das Cruzes. Divisão de Cirurgia Torácica Geral. Mogi das Cruzes. Brasil
  • Canzian, Mauro; Universidade de Mogi das Cruzes. Divisão de Cirurgia Torácica Geral. Mogi das Cruzes. Brasil
  • Correia, Aristides Tadeu; Universidade de Mogi das Cruzes. Divisão de Cirurgia Torácica Geral. Mogi das Cruzes. Brasil
  • Jatene, Fabio Biscegli; Universidade de Mogi das Cruzes. Divisão de Cirurgia Torácica Geral. Mogi das Cruzes. Brasil
Acta cir. bras. ; 28(1): 10-18, 2013. ilus, tab
Article em En | VETINDEX | ID: vti-8930
Biblioteca responsável: BR1.1
Localização: BR68.1
ABSTRACT

PURPOSE:

To investigate the compatibility of a new model of self-expandable tracheal stent in rats.

METHODS:

A new device of polyurethane covered and non - covered stent was placed in the trachea of Wistar rats. Animals were distributed in two groups the polyurethane covered and non-covered group. Macroscopic parameters included position within the tracheal lumen, adherence to the mucosa, degree of dilatation, permeability and internal diameter. Microscopic findings evaluated were incorporation, inflammatory activity, granulation tissue and epithelial revetment injuries. The observation follow-up was six weeks. All parameters were quantified based on determined score values. Incorporation of the stents was evaluated based on the observation if the stent was fixed into the trachea or if it could be removed. Degree of dilatation was performed by external diameter measurements. Granulation tissue was evaluated by measurements of height of the tissue growing into the tracheal lumen.

RESULTS:

100% of non-covered stents had total attachment to mucosa and 100% of polyurethane covered type had adherence only. Regarding dilatation, granulation tissue, inflammatory activity and internal diameter measurements, there were no significant differences between the groups. Pathological tracheal wall injuries were present in both groups.

CONCLUSION:

Both models of stent demonstrated biocompatibility with the trachea. Rats are suitable for an experimental model of tracheal stent study.(AU)
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: VETINDEX Idioma: En Revista: Acta cir. bras. Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: VETINDEX Idioma: En Revista: Acta cir. bras. Ano de publicação: 2013 Tipo de documento: Article