Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 9 de 9
Filtrar
1.
Rep Pract Oncol Radiother ; 19(2): 120-34, 2014 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-24936331

RESUMO

Total skin electron irradiation (TSEI) for patients with cutaneous lymphomas is technically challenging, and numerous approaches have been developed to overcome the many field matching problems associated with such a large and complex treatment volume. Since 1981 we have delivered TSEI using a rotational total skin electron irradiation (RTSEI) technique in conjunction with patch, treat and boost fields in order to provide complete skin and dose coverage. Initially we used a 6 MeV electron beam at an extended source-skin distance (SSD) on a modified linear accelerator. More recently we began using a high dose rate electron mode on a commercially available linear accelerator. The RTSEI technique allows the delivery of a seamless surface dose to the majority of the patient's skin surface in a single treatment. In this review paper we present our three-decade experience with the technical development, dosimetry, treatment delivery and clinical outcomes of our RTSEI technique.

2.
J Appl Clin Med Phys ; 14(6): 4239, 2013 Nov 04.
Artigo em Inglês | MEDLINE | ID: mdl-24257270

RESUMO

Remote afterloading devices used for high-dose-rate (HDR) brachytherapy may be supplied with different sources, and these sources typically have differing initial source strengths. In addition, the proposed frequency for source changes may also vary, depending upon the source type. Dosimetric parameters unique to each source are often used to compare source types. However, when considering the relative dosimetric efficiency between two HDR sources, the combined effect of source type, initial source strength, and source change scheme must be considered. A method of quantifying this combined effect by calculating mean dose rate from specific dosimetric source data is discussed. This method suggests an objective manner of comparing source scheme equivalency to facilitate performing a cost ratio analysis between different HDR sources and source change schemes.


Assuntos
Braquiterapia , Radioisótopos de Cobalto , Radioisótopos de Irídio , Neoplasias/radioterapia , Radiometria , Humanos , Método de Monte Carlo , Prognóstico , Dosagem Radioterapêutica
3.
Int J Radiat Oncol Biol Phys ; 117(3): 763-773, 2023 11 01.
Artigo em Inglês | MEDLINE | ID: mdl-37150259

RESUMO

PURPOSE: The intraoperative radiotherapy in newly diagnosed glioblastoma multiforme (INTRAGO) clinical trial assesses survival in patients with glioblastoma treated with intraoperative radiation therapy (IORT) using the INTRABEAM. Treatment planning for INTRABEAM relies on vendor-provided in-water depth dose curves obtained according to the TARGeted Intraoperative radioTherapy (TARGIT) dosimetry protocol. However, recent studies have shown discrepancies between the estimated TARGIT and delivered doses. This work evaluates the effect of the choice of dosimetry formalism on organs at risk (OAR) doses. METHODS AND MATERIALS: A treatment planning framework for INTRABEAM was developed to retrospectively calculate the IORT dose in 8 INTRAGO patients. These patients received an IORT prescription dose of 20 to 30 Gy in addition to external beam radiation therapy. The IORT dose was obtained using (1) the TARGIT method; (2) the manufacturer's V4.0 method; (3) the CQ method, which uses an ionization chamber Monte Carlo (MC) calculated factor; (4) MC dose-to-water; and (5) MC dose-to-tissue. The IORT dose was converted to 2 Gy fractions equivalent dose. RESULTS: According to the TARGIT method, the OAR dose constraints were respected in all cases. However, the other formalisms estimated a higher mean dose to OARs and revealed 1 case where the constraint for the brain stem was exceeded. The addition of the external beam radiation therapy and TARGIT IORT doses resulted in 10 cases of OARs exceeding the dose constraints. The more accurate MC calculation of dose-to-tissue led to the highest dosimetric differences, with 3, 3, 2, and 2 cases (out of 8) exceeding the dose constraint to the brain stem, optic chiasm, optic nerves, and lenses, respectively. Moreover, the mean cumulative dose to brain stem exceeded its constraint of 66 Gy with the MC dose-to-tissue method, which was not evident with the current INTRAGO clinical practice. CONCLUSIONS: The current clinical approach of calculating the IORT dose with the TARGIT method may considerably underestimate doses to nearby OARs. In practice, OAR dose constraints may have been exceeded, as revealed by more accurate methods.


Assuntos
Neoplasias da Mama , Glioblastoma , Feminino , Humanos , Glioblastoma/radioterapia , Glioblastoma/cirurgia , Órgãos em Risco/diagnóstico por imagem , Órgãos em Risco/efeitos da radiação , Radiometria , Dosagem Radioterapêutica , Estudos Retrospectivos
4.
Phys Med Biol ; 66(21)2021 11 05.
Artigo em Inglês | MEDLINE | ID: mdl-34663769

RESUMO

OBJECTIVE: The relative TG-43 dosimetry parameters of the INTRABEAM (Carl Zeiss Meditec AG, Jena, Germany) bare probe were recently reported by Ayala Alvarezet al(2020Phys. Med. Biol.65245041). The current study focuses on the dosimetry characterization of the INTRABEAM source with the eight available spherical applicators according to the TG-43 formalism using Monte Carlo (MC) simulations. APPROACH: This report includes the calculated dose-rate conversion coefficients that determine the absolute dose rate to water at a reference point of 10 mm from the applicator surface, based on calibration air-kerma rate measurements at 50 cm from the source on its transverse plane. Since the air-kerma rate measurements are not yet provided from a standards laboratory for the INTRABEAM, the values in the present study were calculated with MC. This approach is aligned with other works in the search for standardization of the dosimetry of electronic brachytherapy sources. As a validation of the MC model, depth dose calculations along the source axis were compared with calibration data from the source manufacturer. MAIN RESULTS: The calculated dose-rate conversion coefficients were 434.0 for the bare probe, and 683.5, 548.3, 449.9, 376.5, 251.0, 225.6, 202.8, and 182.6 for the source with applicators of increasing diameter from 15 to 50 mm, respectively. The radial dose and the 2D anisotropy functions of the TG-43 formalism were also obtained and tabulated in this document. SIGNIFICANCE: This work presents the data required by a treatment planning system for the characterization of the INTRABEAM system in the context of intraoperative radiotherapy applications.


Assuntos
Braquiterapia , Radiometria , Calibragem , Método de Monte Carlo , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
5.
Phys Med Biol ; 65(24): 245041, 2020 12 22.
Artigo em Inglês | MEDLINE | ID: mdl-33137796

RESUMO

The INTRABEAM system (Carl Zeiss Meditec AG, Jena, Germany) is an electronic brachytherapy (eBT) device designed for intraoperative radiotherapy applications. To date, the INTRABEAM x-ray source has not been characterized according to the AAPM TG-43 specifications for brachytherapy sources. This restricts its modelling in commercial treatment planning systems (TPSs), with the consequence that the doses to organs at risk are unknown. The aim of this work is to characterize the INTRABEAM source according to the TG-43 brachytherapy dosimetry protocol. The dose distribution in water around the source was determined with Monte Carlo (MC) calculations. For the validation of the MC model, depth dose calculations along the source longitudinal axis were compared with measurements using a soft x-ray ionization chamber (PTW 34013) and two synthetic diamond detectors (microDiamond PTW TN60019). In our results, the measurements in water agreed with the MC model calculations within uncertainties. The use of the microDiamond detector yielded better agreement with MC calculations, within estimated uncertainties, compared to the ionization chamber at points of steeper dose gradients. The radial dose function showed a steep fall-off close to the INTRABEAM source ([Formula: see text]10 mm) with a gradient higher than that of commonly used brachytherapy radionuclides (192Ir, 125I and 103Pd), with values of 2.510, 1.645 and 1.232 at 4, 6 and 8 mm, respectively. The radial dose function partially flattens at larger distances with a fall-off comparable to that of the Xoft Axxent® (iCAD, Inc., Nashua, NH) eBT system. The simulated 2D polar anisotropy close to the bare probe walls showed deviations from unity of up to 55% at 10 mm and 155°. This work presents the MC calculated TG-43 parameters for the INTRABEAM, which constitute the necessary data for the characterization of the source as required by a TPS used in clinical dose calculations.


Assuntos
Braquiterapia , Método de Monte Carlo , Radiometria , Anisotropia , Humanos , Período Intraoperatório , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Incerteza
6.
J Appl Clin Med Phys ; 9(4): 123-134, 2008 Nov 03.
Artigo em Inglês | MEDLINE | ID: mdl-19020483

RESUMO

The rotational total skin electron irradiation (RTSEI) technique at our institution has undergone several developments over the past few years. Replacement of the formerly used linear accelerator has prompted many modifications to the previous technique. With the current technique, the patient is treated with a single large field while standing on a rotating platform, at a source-to-surface distance of 380 cm. The electron field is produced by a Varian 21EX linear accelerator using the commercially available 6 MeV high dose rate total skin electron mode, along with a custom-built flattening filter. Ionization chambers, radiochromic film, and MOSFET (metal oxide semiconductor field effect transistor) detectors have been used to determine the dosimetric properties of this technique. Measurements investigating the stationary beam properties, the effects of full rotation, and the dose distributions to a humanoid phantom are reported. The current treatment technique and dose regimen are also described.


Assuntos
Aceleradores de Partículas/instrumentação , Radioterapia/métodos , Pele/efeitos da radiação , Calibragem , Computadores , Relação Dose-Resposta à Radiação , Elétrons , Desenho de Equipamento , Humanos , Íons , Óxidos/química , Imagens de Fantasmas , Semicondutores , Pele/patologia , Filme para Raios X
7.
Phys Med Biol ; 63(17): 175007, 2018 08 30.
Artigo em Inglês | MEDLINE | ID: mdl-30095077

RESUMO

Despite being considered the gold standard for brachytherapy dosimetry, Monte Carlo (MC) has yet to be implemented into a software for brachytherapy treatment planning. The purpose of this work is to present RapidBrachyMCTPS, a novel treatment planning system (TPS) for brachytherapy applications equipped with a graphical user interface (GUI), optimization tools and a Geant4-based MC dose calculation engine, RapidBrachyMC. Brachytherapy sources and applicators were implemented in RapidBrachyMC and made available to the user via a source and applicator library in the GUI. To benchmark RapidBrachyMC, TG-43 parameters were calculated for the microSelectron v2 (192Ir) and SelectSeed (125I) source models and were compared against previously validated MC brachytherapy codes. The performance of RapidBrachyMC was evaluated for a prostate high dose rate case. To assess the accuracy of RapidBrachyMC in a heterogeneous setup, dose distributions with a cylindrical shielded/unshielded applicator were validated against film measurements in a Solid WaterTM phantom. TG-43 parameters calculated using RapidBrachyMC generally agreed within 1%-2% of the results obtained in previously published work. For the prostate case, clinical dosimetric indices showed general agreement with Oncentra TPS within 1%. Simulation times were on the order of minutes on a single core to achieve uncertainties below 2% in voxels within the prostate. The calculation time was decreased further using the multithreading features of Geant4. In the comparison between MC-calculated and film-measured dose distributions, at least 95% of points passed the 3%/3 mm gamma index criteria in all but one case. RapidBrachyMCTPS can be used as a post-implant dosimetry toolkit, as well as for MC-based brachytherapy treatment planning. This software is especially well suited for the development of new source and applicator models.


Assuntos
Braquiterapia/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Software , Humanos , Método de Monte Carlo , Imagens de Fantasmas , Dosagem Radioterapêutica
8.
J Appl Clin Med Phys ; 7(1): 42-51, 2006.
Artigo em Inglês | MEDLINE | ID: mdl-16518316

RESUMO

While the current trend in radiotherapy is to replace cobalt teletherapy units with more versatile and technologically advanced linear accelerators, there remain some useful applications for older cobalt units. The expansion of our radiotherapy department involved the decommissioning of an isocentric cobalt teletherapy unit and the replacement of a column-mounted 4-MV LINAC that has been used for total body irradiation (TBI). To continue offering TBI treatments, we converted the decommissioned cobalt unit into a dedicated fixed-field total body irradiator and installed it in an existing medium-energy LINAC bunker. This article describes the logistical and dosimetric aspects of bringing a reconditioned cobalt teletherapy unit into clinical service as a total body irradiator.


Assuntos
Radioisótopos de Cobalto/uso terapêutico , Teleterapia por Radioisótopo/instrumentação , Irradiação Corporal Total/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Reutilização de Equipamento
9.
Med Phys ; 32(9): 2899-910, 2005 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-16266104

RESUMO

A study of the induced activity in a medical linear accelerator (linac) room was carried out on several linac installations. Higher beam energy, higher dose rate, and larger field size generally result in higher activation levels at a given point of interest, while the use of multileaf collimators (MLC) can also increase the activation level at the isocenter. Both theoretical and experimental studies reveal that the activation level in the morning before any clinical work increases from Monday to Saturday and then decreases during the weekend. This weekly activation picture keeps stable from one week to another during standard clinical operation of the linac. An effective half-life for a given point in the treatment room can be determined from the measured or calculated activity decay curves. The effective half-life for points inside the treatment field is longer than that for points outside of the field in the patient plane, while a larger field and longer irradiation time can also make the effective half-life longer. The activation level reaches its practical saturation value after a 30 min continuous irradiation, corresponding to 12 000 MU at a "dose rate" of 400 MU/min. A "dose" of 300 MU was given 20 times in 15 min intervals to determine the trends in the activation level in a typical clinical mode. As well, a long-term (85 h over a long weekend) decay curve was measured to evaluate the long-term decay of room activation after a typical day of clinical linac use. A mathematical model for the activation level at the isocenter has been established and shown to be useful in explaining and predicting the induced activity levels for typical clinical and experimental conditions. The activation level for a 22 MeV electron beam was also measured and the result shows it is essentially negligible.


Assuntos
Modelos Teóricos , Aceleradores de Partículas , Monitoramento de Radiação , Dosagem Radioterapêutica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA