Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 9 de 9
Filtrar
Mais filtros

Base de dados
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 109: 425-429, 2016 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-26651164

RESUMO

In this new approach, the active shielding of a low level spectrometer is monitored by an extendable dead time used as an "extendable GATE" signal. This concept is directly inspired from the live-timed anti-coincidence system implemented at LNE-LNHB (MTR2 modules). This allows a significant reduction in the cost and complexity of the system since several electronic modules are replaced by only one module dedicated specifically to this type of experiment. Settings of this new approach are detailed and results are discussed.

2.
Appl Radiat Isot ; 114: 167-72, 2016 08.
Artigo em Inglês | MEDLINE | ID: mdl-27236833

RESUMO

Two series of activity standards of (60)Co in cast steel matrix, developed for the calibration of gamma-ray spectrometry systems in the metallurgical sector, were characterised using a European interlaboratory comparison among twelve National Metrology Institutes and one international organisation. The first standard, consisting of 14 disc shaped samples, was cast from steel contaminated during production ("originally"), and the second, consisting of 15 similar discs, from artificially-contaminated ("spiked") steel. The reference activity concentrations of (60)Co in the cast steel standards were (1.077±0.019) Bqg(-1) on 1 January 2013 12h00 UT and (1.483±0.022) Bqg(-1) on 1 June 2013 12h00 UT, respectively.

3.
Appl Radiat Isot ; 60(2-4): 159-65, 2004.
Artigo em Inglês | MEDLINE | ID: mdl-14987636

RESUMO

A tunable monochromatic X-ray source operating in the 1-20 keV energy range is described. An X-ray tube provides initial photons. A dispersive crystal performs the energy selection, according to Bragg's law. An X-ray detector is connected to the monochromator fixed exit. This setup can be used for metrological studies. A first application consists in measuring attenuation coefficients in the 4-10 keV energy range. Results for aluminum and copper are given, with average relative uncertainties (1sigma) of 1% and 3% respectively.


Assuntos
Calibragem/normas , Análise de Falha de Equipamento , Radiometria/instrumentação , Radiometria/normas , Padrões de Referência , Transdutores , Raios X , Desenho de Equipamento , Doses de Radiação , Radiometria/métodos , Reprodutibilidade dos Testes , Espalhamento de Radiação , Sensibilidade e Especificidade
4.
Appl Radiat Isot ; 81: 71-5, 2013 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-23615361

RESUMO

In this study, different aspects of the Compton scattering inside volume sources are illustrated using experimental approach and Monte Carlo simulation. For the low-energy range (below 100 keV) scattered events represents around 30% of the whole spectrum. Influence of the source-detector geometry is discussed. The scattering effects induce strong differences in spectrum shape for different geometrical conditions. This should influence efficiency transfer factors. A new approach is proposed, including the scattered events, to avoid complex peak area determination.


Assuntos
Modelos Estatísticos , Método de Monte Carlo , Radiometria/métodos , Software , Análise Espectral/métodos , Simulação por Computador , Transferência de Energia , Fótons , Espalhamento de Radiação
5.
Appl Radiat Isot ; 81: 67-70, 2013 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-23623315

RESUMO

In the framework of environmental measurements by gamma-ray spectrometry, some laboratories need to characterize samples in geometries for which a calibration is not directly available. A possibility is to use an efficiency transfer code, e.g., ETNA. However, validation for large volume sources, such as Marinelli geometries, is needed. With this aim in mind, ETNA is compared, initially to a Monte Carlo simulation (PENELOPE) and subsequently to experimental data obtained with a high-purity germanium detector (HPGe).


Assuntos
Algoritmos , Modelos Estatísticos , Monitoramento de Radiação/métodos , Monitoramento de Radiação/normas , Software , Espectrometria gama/métodos , Espectrometria gama/normas , Simulação por Computador , Internacionalidade , Transferência Linear de Energia , Doses de Radiação
6.
Appl Radiat Isot ; 81: 114-8, 2013 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-23582499

RESUMO

In this work, PENELOPE2008 and MCNP5 software packages were used to investigate the effect of simulation parameters such as cut-off energy of photons and electrons on the detector response function and peak escape intensity to assess their influence on the efficiency determination. The results showed a general agreement between the two codes, however both the intensity of escape peaks and total efficiency computed with PENELOPE2008 are higher than those obtained with MCNP5. Moreover, the calculated efficiencies are typically 8% higher than what is experimentally found in the low energy range. Further investigation was required to obtain better agreement and lead to better estimate of the real germanium crystal dimensions and dead layer.


Assuntos
Algoritmos , Artefatos , Radiação Cósmica , Modelos Estatísticos , Método de Monte Carlo , Software , Espectrometria gama/instrumentação , Simulação por Computador , Desenho Assistido por Computador , Desenho de Equipamento , Análise de Falha de Equipamento , Doses de Radiação
7.
Appl Radiat Isot ; 70(9): 2137-40, 2012 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-22410298

RESUMO

The numerical calculation of coincidence summing corrections requires both the total and full-energy peak efficiencies to be included in the corrective factors. Moreover, in the case of volume sources, the coincidence probability depends on the position of the photon emission inside the radioactive sample, thus it is necessary to calculate the coincidence summing corrective factors by integrating the elemental contributions for the entire volume. The software ETNA calculates coincidence summing corrective factors according to a deterministic method, for volume sources, two approaches are included in the software: the "simplified" calculation uses directly the efficiency of the volume source like in the case of point source, and the "complete" calculation performs the volume integration. To assess the difference between these, different geometrical cases are tested for several radionuclides using both methods.


Assuntos
Algoritmos , Artefatos , Modelos Estatísticos , Radioisótopos/análise , Radiometria/métodos , Software , Simulação por Computador , Doses de Radiação
8.
Appl Radiat Isot ; 68(7-8): 1423-7, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-20022512

RESUMO

The activity measurement of beta-plus emitters by gamma-ray spectrometry is studied. Experimental measurements are performed with (22)Na, (65)Zn and (64)Cu with sources included in a lead container. For these nuclides, the activity can be derived both from one photon emission peak and from the 511 keV annihilation peak, including annihilation in-flight correction and geometry correction computed by Monte Carlo simulation. The activity values obtained using the two types of peaks show satisfying agreement. The extension of the method to volume sources is discussed.


Assuntos
Partículas beta , Radioisótopos/análise , Espectrometria gama/métodos , Cobre/análise , Métodos , Sódio/análise , Zinco/análise
9.
Appl Radiat Isot ; 67(5): 957-60, 2009 May.
Artigo em Inglês | MEDLINE | ID: mdl-19285421

RESUMO

Gamma-ray spectrometry measurements have been carried out to determine the activity of natural radionuclides in a phosphogypsum sample included in a specific tight container. The gamma spectrometer includes an N-type coaxial high-purity germanium (HPGe) detector equipped with an anti-cosmic system. This measurement required the determination of linear attenuation coefficients of phosphogypsum to calculate self-absorption correction between efficiency calibration conditions and measurement ones. The results are given for the three natural chains and for (40)K, in term of specific activity/g of dry material, ranging from a few Bq kg(-1) to a few hundreds Bq kg(-1). The equilibrium within the different families and the (235)U/(238)U ratio are discussed.


Assuntos
Sulfato de Cálcio/análise , Fósforo/análise , Radioisótopos/análise , Espectrometria gama/métodos , Absorção , Calibragem , Espectrometria gama/instrumentação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA