Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Mais filtros

Base de dados
Tipo de documento
Intervalo de ano de publicação
2.
Z Kardiol ; 86(7): 541-4, 1997 Jul.
Artigo em Alemão | MEDLINE | ID: mdl-9340946

RESUMO

This report documents a case of ball variance in a Smeloff-Cutter aortic prosthesis occurring 24 years after implantation. After episodes of embolic complications the patient died in acute shock. The silicone rubber ball showed several alterations including discoloration, grooving, cracking, swelling and subtotal fracture of the poppet. Terminal valvular malfunction was caused by complete thrombosis of the prosthesis. In most patients ball variance occurred during the first years after valve replacement; thus, the observed case is a very rare late complication of a ball-valve prosthesis.


Assuntos
Insuficiência da Valva Aórtica/cirurgia , Análise de Falha de Equipamento , Próteses Valvulares Cardíacas , Complicações Pós-Operatórias/patologia , Valva Aórtica/patologia , Insuficiência da Valva Aórtica/patologia , Evolução Fatal , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Trombose/patologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA