RESUMO
OBJECTIVE: To report the investigational device exemption study 1-year clinical outcomes of the high neck angulation (HNA) substudy of the GORE EXCLUDER Conformable AAA Endoprosthesis (EXCC) for treatment of infrarenal abdominal aortic aneurysms (AAAs). METHODS: This study is a prospective, multicenter clinical trial conducted in the United States and included core laboratory assessment of imaging and independent event adjudication. Anatomical criteria for enrollment in the HNA substudy included infrarenal aortic neck angulation >60° and ≤90° with aortic neck length ≥10 mm. Primary safety end points included estimated blood loss of >1000 mL, death, stroke, myocardial infarction, bowel ischemia, paraplegia, respiratory failure, renal failure, and thromboembolic events. Primary effectiveness end points included technical success, absence from type I and III endoleaks, migration (≥10 mm), sac enlargement (≥5 mm), sac rupture, and conversion to open repair. RESULTS: Between January 2018 and February 2022, 95 patients were enrolled in the HNA substudy across 35 sites. Of the 95 patients, 71 (74.7%) were male and the cohort average age was 74.4 years. The mean infrarenal proximal aortic neck angle was 71.6° and the mean AAA size was 62.9 mm. Overall technical success was achieved in 93 patients (97.9%). Freedom from a primary safety end point through 30 days was 96.7%; 3 (3.3%) patients had an estimated blood loss of >1000 mL. Freedom from the primary effectiveness at 12 months was achieved in 94.8%. Four patients (4.3%) had a type IA endoleak; intervention after the procedure was not required and no subsequent interventions or sac enlargement were noted in these patients. At 12 months, 29 patients (39.7%) experienced a type II endoleak and 1 (1.3%) patient experienced AAA sac expansion of ≥5 mm. Through 12 months, 1 patient (1.3%) had a conversion to open surgical repair. There were no aneurysm-related deaths, ruptures, or migration through 12 months. CONCLUSIONS: The investigational device exemption study demonstrates safety and effectiveness of the GORE EXCLUDER Conformable AAA Endoprosthesis device in AAA with highly angulated necks (>60° and ≤90°) are preserved at the 12-month follow-up.
RESUMO
OBJECTIVE: To report the midterm clinical outcomes from the GORE® EXCLUDER® Conformable AAA Endoprosthesis system (EXCC) pivotal regulatory trial in the United States (U.S.). METHODS: This is a prospective, multicenter, investigational device exemption clinical trial at 31 U S. sites with Core Laboratory assessment of imaging and independent adjudication of safety. The study enrolled patients with abdominal aortic aneurysms (AAA) with a minimum proximal landing zone ≥10 mm and proximal neck angulation of ≤60 degrees between December 2017 and February 2019 as part of a larger study to gain indications of the EXCC device. Endpoints included patient survival, freedom from secondary interventions, and stent-graft related outcomes. RESULTS: There were 80 patients enrolled (88.8% male, mean 73.5 ± 8.14 years-old). Mean maximum aortic diameter was 57.7±8.0 mm (range, 42.5-82.7). There was 100% freedom from type I and III endoleak and aneurysm-related mortality at 36-months. Freedom from secondary intervention was 91.9 ± (0.83, 0.96, 95% C.I.) at 36-months. There were no device fractures, migrations (≥10 mm), or aneurysm ruptures. At 36 months, thirteen patients (26.5%) had type 2 endoleak, 32 patients (58.2%) had AAA sac regression, 17 (30.9%) had no change in diameter, and 6 (10.9%) had sac enlargement. Seven patients (8.8%) through 36 months underwent reintervention. CONCLUSIONS: The 3-year outcomes have continued to show an adequate safety and efficacy profile of the EXCC device with no aneurysm related mortality or Type I/III endoleak. These results demonstrate durability for an EVAR device in US regulatory trials.
RESUMO
BACKGROUND: This study reports the 30-day outcomes of the primary arm of the GORE EXCLUDER Thoracoabdominal Branch Endoprosthesis (TAMBE) pivotal trial for complex abdominal aortic aneurysm repair. METHODS: This multicenter, nonrandomized, prospective study of the TAMBE device included patients enrolled in the primary study arm of extent IV thoracoabdominal aortic aneurysms and pararenal aneurysms. Technical success and major adverse events were analyzed per the Society for Vascular Surgery guidelines. RESULTS: The 102 patients of the primary arm who underwent endovascular repair using the TAMBE device were a mean age of 73 ± 6.4 years (range, 58-82 years) and 84 (84.2%) were male. The mean body mass index was 28.3 ± 5.0 kg/m2. Fifty-nine patients (57.8%) were treated for extent IV and 43 (42.2%) pararenal aneurysms; the mean maximum diameter of the aneurysms was 59.4 ± 7.8 mm. A prophylactic cerebral spinal fluid drain was used in 10 patients (9.8%). Technical success was achieved in 99% of patients, with the single failure owing to unsuccessful cannulation of the left renal artery. Mean procedure time was 315 ± 103 minutes (range, 163-944 minutes), estimated blood loss was 300 ± 296 mL (range, 10-2000 mL), and contrast administration was 153.6 ± 73.5 mL (range, 16-420 mL). The intensive care unit length of stay was 58.7 ± 52.7 hours (range, 1-288 mL). In 28 patients (27.5%), a total of 32 additional endovascular components were deployed to manage procedural complications including aortic and target vessel dissections and injuries not related to access. Bridging stent grafts were deployed to incorporate 407 target vessels (mean 1.6/per vessel; range, 1-4). Postoperative transfusion was required in 14 patients (13.7%). Major adverse events occurred in seven patients (6.9%) through 30 days. Events included respiratory failure (n = 2), disabling stroke (n = 1), new-onset renal failure requiring dialysis (n = 2), and paraplegia (n = 2). At 30 days, there was one patient with intraoperative rupture; no severe bowel ischemia or lesion-related/all-cause mortality were reported. The Core lab-reported patency was 100% in the aortic component, superior mesenteric artery, and celiac artery, and 95.9% in the left renal and 99.0% in the right renal branch components through 30 days of follow-up. Reinterventions through 30 days were performed in 9 of 96 patients (9.4%) and were all minor. CONCLUSIONS: Early TAMBE device outcomes demonstrate a high technical success rate, no 30-day lesion-related mortality, and a low rate of safety events within 30 days of the index procedure.
Assuntos
Aneurisma da Aorta Abdominal , Aneurisma da Aorta Torácica , Implante de Prótese Vascular , Prótese Vascular , Procedimentos Endovasculares , Complicações Pós-Operatórias , Desenho de Prótese , Humanos , Masculino , Idoso , Feminino , Aneurisma da Aorta Torácica/cirurgia , Aneurisma da Aorta Torácica/mortalidade , Aneurisma da Aorta Torácica/diagnóstico por imagem , Pessoa de Meia-Idade , Idoso de 80 Anos ou mais , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/instrumentação , Procedimentos Endovasculares/mortalidade , Estudos Prospectivos , Resultado do Tratamento , Implante de Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/instrumentação , Implante de Prótese Vascular/mortalidade , Aneurisma da Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/mortalidade , Fatores de Tempo , Complicações Pós-Operatórias/etiologia , Fatores de Risco , Stents , Estados Unidos , Tempo de Internação , Aneurisma da Aorta ToracoabdominalRESUMO
OBJECTIVE: Despite the publication of various national/international guidelines, several questions concerning the management of patients with asymptomatic (AsxCS) and symptomatic (SxCS) carotid stenosis remain unanswered. The aim of this international, multi-specialty, expert-based Delphi Consensus document was to address these issues to help clinicians make decisions when guidelines are unclear. METHODS: Fourteen controversial topics were identified. A three-round Delphi Consensus process was performed including 61 experts. The aim of Round 1 was to investigate the differing views and opinions regarding these unresolved topics. In Round 2, clarifications were asked from each participant. In Round 3, the questionnaire was resent to all participants for their final vote. Consensus was reached when ≥75% of experts agreed on a specific response. RESULTS: Most experts agreed that: (1) the current periprocedural/in-hospital stroke/death thresholds for performing a carotid intervention should be lowered from 6% to 4% in patients with SxCS and from 3% to 2% in patients with AsxCS; (2) the time threshold for a patient being considered "recently symptomatic" should be reduced from the current definition of "6 months" to 3 months or less; (3) 80% to 99% AsxCS carries a higher risk of stroke compared with 60% to 79% AsxCS; (4) factors beyond the grade of stenosis and symptoms should be added to the indications for revascularization in AsxCS patients (eg, plaque features of vulnerability and silent infarctions on brain computed tomography scans); and (5) shunting should be used selectively, rather than always or never. Consensus could not be reached on the remaining topics due to conflicting, inadequate, or controversial evidence. CONCLUSIONS: The present international, multi-specialty expert-based Delphi Consensus document attempted to provide responses to several unanswered/unresolved issues. However, consensus could not be achieved on some topics, highlighting areas requiring future research.
Assuntos
Estenose das Carótidas , Acidente Vascular Cerebral , Humanos , Estenose das Carótidas/diagnóstico , Estenose das Carótidas/diagnóstico por imagem , Consenso , Técnica Delphi , Acidente Vascular Cerebral/diagnóstico , Acidente Vascular Cerebral/etiologia , Constrição PatológicaRESUMO
BACKGROUND: Since the risk of mortality from rupture is elevated, elective repair of abdominal aortic aneurysms (AAAs) is often recommended. Currently, over 80% of elective repairs are carried out using an endovascular approach. While open repair has similar late survival and fewer reintervention outcomes when compared to endovascular repair, incisional hernia is a frequent complication with morbidity and cost implications. The Open versus Endovascular Repair (OVER) trial was the largest randomized trial of endovascular versus open repair of AAA in the United States. The purpose of this study was to determine risk factors associated with incisional hernia development following AAA repair via secondary analysis of the OVER data. METHODS: This was a multisite trial conducted within the Veterans Affairs health-care system. Study participants (N = 881) were enrolled from 2002 to 2008 and followed until 2011 with additional administrative data collection until 2016. Eligible patients had AAA for which elective repair was planned and randomized 1:1 to either open or endovascular repair. Incisional hernia was a prespecified end point in the OVER protocol, specifically assessed at each protocol follow-up visit. Technical details were extracted from each operative report, repair case report form(s), and adverse event form(s). Patient demographics, comorbid conditions, reported preoperative activity level, and operative details including initial approach, blood loss, and closure methods were analyzed using Bayesian hierarchical Weibull survival regression modeling. RESULTS: Incisional hernias were recorded among 46 participants (5.2%). The average time to hernia diagnosis was 3.5 years. Of the 437 participants randomized to open treatment, 427 received an open repair including crossovers from endovascular treatment assignment. Transperitoneal repair was performed in 81%, running suture in 96%, and absorbable suture in 71% of cases. Randomization to endovascular repair was associated with reduced risk of hernia (hazard ratio [HR] 0.70, 95% credible interval [CI] 0.49-0.94). Higher activity level was associated with increased hernia risk (HR 1.39, 95% CI 1.06-1.84). Approach, suture closure techniques, body mass index, diabetes, and smoking status were not associated with increased risk of hernia development. CONCLUSIONS: Incisional hernia is a frequent complication associated with open repair of abdominal aortic aneurysm and commonly required reintervention. Endovascular repair was associated with reduced risk of hernia. Patients with increased activity experienced a higher incidence of hernia. However, no other modifiable patient, operative, or technical factors were found to be associated with hernia development.
Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Procedimentos Endovasculares , Hérnia Incisional , Humanos , Fatores de Risco , Masculino , Aneurisma da Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/mortalidade , Idoso , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/mortalidade , Feminino , Hérnia Incisional/cirurgia , Hérnia Incisional/etiologia , Resultado do Tratamento , Fatores de Tempo , Implante de Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/mortalidade , Medição de Risco , Estados Unidos/epidemiologia , Pessoa de Meia-Idade , Idoso de 80 Anos ou maisRESUMO
BACKGROUND: Endovascular repair of blunt thoracic aortic injury (BTAI) has dramatically reduced the morbidity and mortality of intervention. Injuries requiring zone 2 coverage of the aorta traditionally require left subclavian artery (LSA) sacrifice or open revascularization. Furthermore, these injuries are associated with an increased risk of in-hospital mortality and long-term morbidity. Here we report 1-year outcomes of total endovascular repair of BTAI with the GORE® TAG® Thoracic Branch Endoprosthesis for LSA preservation. METHODS: Across 34 investigative sites, 9 patients with BTAI requiring LSA coverage were enrolled in a nonrandomized, prospective study of a single-branched aortic endograft. The thoracic branch endoprosthesis device allows for graft placement proximal to the LSA and incorporates a single side branch for LSA perfusion. RESULTS: This initial cohort included 8 male and 1 female patient with a median age of 43 (22, 76) and 12 months of follow-up. Five total years of follow-up are planned. All participants had grade 3 BTAI. All procedures took place between 2018 and 2019. The median injury severity score was 2 (0, 66). The median procedure time was 109 min (78, 162). All aortic injuries were repaired under general anesthesia and with heparinization. A spinal drain was used in one patient. Post-deployment balloon angioplasty was conducted in one case at the distal landing zone. There was one asymptomatic LSA branch occlusion 6 months after repair. It was attributed to the purposeful proximal deployment of the branch stent to accommodate an early vertebral takeoff. The occlusion did not require revascularization. There were no strokes, mortalities, or aortic adverse events (migration, endoleak, native aortic expansion, dissection, or thrombosis) through 12 months of follow-up. CONCLUSIONS: Initial cohort outcomes suggest that endovascular repair of zone 2 BTAI is feasible and has favorable outcomes using the thoracic branch device with LSA preservation. Additional cases and longer-term follow-up are required for a definitive assessment of the device's safety and durability in traumatic aortic injuries.
Assuntos
Aorta Torácica , Implante de Prótese Vascular , Prótese Vascular , Procedimentos Endovasculares , Desenho de Prótese , Lesões do Sistema Vascular , Ferimentos não Penetrantes , Humanos , Masculino , Ferimentos não Penetrantes/cirurgia , Ferimentos não Penetrantes/diagnóstico por imagem , Feminino , Lesões do Sistema Vascular/cirurgia , Lesões do Sistema Vascular/diagnóstico por imagem , Lesões do Sistema Vascular/mortalidade , Lesões do Sistema Vascular/etiologia , Implante de Prótese Vascular/instrumentação , Implante de Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/mortalidade , Aorta Torácica/cirurgia , Aorta Torácica/diagnóstico por imagem , Aorta Torácica/lesões , Procedimentos Endovasculares/instrumentação , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/mortalidade , Adulto , Resultado do Tratamento , Pessoa de Meia-Idade , Fatores de Tempo , Estudos Prospectivos , Adulto Jovem , Idoso , Traumatismos Torácicos/cirurgia , Traumatismos Torácicos/diagnóstico por imagem , Traumatismos Torácicos/mortalidade , Estados Unidos , Stents , Fatores de RiscoRESUMO
BACKGROUND: The Global Iliac Branch Study (NCT05607277) is an international, multicenter, retrospective cohort study of anatomic predictors of adverse iliac events (AIEs) in aortoiliac aneurysms treated with iliac branch devices (IBDs). METHODS: Patients with pre-IBD and post-IBD computed tomography imaging were included. We measured arterial diameters, stenosis, calcification, bifurcation angles, and tortuosity indices using a standardized, validated protocol. A composite of ipsilateral AIE was defined, a priori, as occlusion, type I or III endoleak, device constriction, or clinical event requiring reintervention. Paired t-test compared tortuosity indices and splay angles pretreatment and post-treatment for all IBDs and by device material (stainless steel and nitinol). Two-sample t-test compared anatomical changes from pretreatment to post-treatment by device material. Logistic regression assessed associations between AIE and anatomic measurements. Analysis was performed by IBD. RESULTS: We analyzed 297 patients (286 males, 11 females) with 331 IBDs (227 stainless steel, 104 nitinol). Median clinical follow-up was 3.8 years. Iliac anatomy was significantly straightened with all IBD treatment, though stainless steel IBDs had a greater reduction in total iliac artery tortuosity index and aortic splay angle compared to nitinol IBDs (absolute reduction -0.20 [-0.22 to -0.18] vs. -0.09 [-0.12 to -0.06], P < 0.0001 and -19.6° [-22.4° to -16.9°] vs. -11.2° [-15.3° to -7.0°], P = 0.001, respectively). There were 54 AIEs in 44 IBDs in 42 patients (AIE in 13.3% of IBD systems), requiring 35 reinterventions (median time to event 41 days; median time to reintervention 153 days). There were 18 endoleaks, 29 occlusions, and 5 device constrictions. There were no strong associations between anatomic measurements and AIE overall, though internal iliac diameter was inversely associated with AIE in nitinol devices (nAIE, nitinol = 8). CONCLUSIONS: Purpose-built IBDs effectively treat aortoiliac disease, including that with tortuous anatomy, with a high patency rate (91.5%) and low reintervention rate (9.1%) at 4 years. Anatomic predictors of AIE are limited.
Assuntos
Ligas , Implante de Prótese Vascular , Prótese Vascular , Angiografia por Tomografia Computadorizada , Procedimentos Endovasculares , Aneurisma Ilíaco , Valor Preditivo dos Testes , Desenho de Prótese , Stents , Humanos , Feminino , Masculino , Aneurisma Ilíaco/diagnóstico por imagem , Aneurisma Ilíaco/cirurgia , Aneurisma Ilíaco/fisiopatologia , Estudos Retrospectivos , Procedimentos Endovasculares/instrumentação , Procedimentos Endovasculares/efeitos adversos , Idoso , Resultado do Tratamento , Implante de Prótese Vascular/instrumentação , Implante de Prótese Vascular/efeitos adversos , Fatores de Tempo , Fatores de Risco , Idoso de 80 Anos ou mais , Pessoa de Meia-Idade , Aortografia , Aço Inoxidável , Endoleak/etiologia , Endoleak/diagnóstico por imagem , Aneurisma Aórtico/diagnóstico por imagem , Aneurisma Aórtico/cirurgia , Aneurisma Aórtico/fisiopatologia , Artéria Ilíaca/diagnóstico por imagem , Artéria Ilíaca/fisiopatologia , Artéria Ilíaca/cirurgia , Estados UnidosRESUMO
OBJECTIVE: Endovascular repair is the preferred treatment for aortoiliac aneurysm, with preservation of at least one internal iliac artery recommended. This study aimed to assess pre-endovascular repair anatomical characteristics of aortoiliac aneurysm in patients from the Global Iliac Branch Study (GIBS, NCT05607277) to enhance selection criteria for iliac branch devices (IBD) and improve long-term outcomes. METHODS: Pre-treatment CT scans of 297 GIBS patients undergoing endovascular aneurysm repair were analyzed. Measurements included total iliac artery length, common iliac artery length, tortuosity index, common iliac artery splay angle, internal iliac artery stenosis, calcification score, and diameters in the device's landing zone. Statistical tests assessed differences in anatomical measurements and IBD-mediated internal iliac artery preservation. RESULTS: Left total iliac artery length was shorter than right (6.7 mm, P = .0019); right common iliac artery less tortuous (P = .0145). Males exhibited greater tortuosity in the left total iliac artery (P = .0475) and larger diameter in left internal iliac artery's landing zone (P = .0453). Preservation was more common on right (158 unilateral, 34 bilateral) than left (105 unilateral, 34 bilateral). There were 192 right-sided and 139 left-sided IBDs, with 318 IBDs in males and 13 in females. CONCLUSION: This study provides comprehensive pre-treatment iliac anatomy analysis in patients undergoing endovascular repair with IBDs, highlighting differences between sides and sexes. These findings could refine patient selection for IBD placement, potentially enhancing outcomes in aortoiliac aneurysm treatment. However, the limited number of females in the study underscores the need for further research to generalize findings across genders.
Assuntos
Procedimentos Endovasculares , Aneurisma Ilíaco , Humanos , Masculino , Feminino , Aneurisma Ilíaco/cirurgia , Aneurisma Ilíaco/diagnóstico por imagem , Idoso , Procedimentos Endovasculares/métodos , Pessoa de Meia-Idade , Aneurisma da Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/patologia , Artéria Ilíaca/diagnóstico por imagem , Implante de Prótese Vascular/métodos , Idoso de 80 Anos ou mais , Tomografia Computadorizada por Raios X , Resultado do TratamentoRESUMO
OBJECTIVES: To determine if sex was an effect modifier in a pooled analysis of asymptomatic patients from CREST and ACT I. MATERIALS AND METHODS: We analyzed data from 2544 patients aged <80 with ≥70 % asymptomatic carotid stenosis randomized to CAS or CEA (nCREST = 1091; nACT-1 = 1453). The pre-specified primary endpoint in both trials was any stroke, myocardial infarction or death during the peri-procedural period, or ipsilateral stroke within 4 years of randomization. The secondary endpoint was any stroke or death during the peri-procedural period, or ipsilateral stroke within 4 years of randomization. RESULTS: There was no significant difference in the frequency of events for men or women between CAS and CEA for the primary or secondary endpoints. When assessing for an interaction of sex and risks between procedures, the treatment-by-sex interaction was not significant for either primary or secondary endpoints in the four-year period or the peri-procedural period. However, women had significantly fewer post-procedural events (ipsilateral stroke) with CAS than CEA (HR = 0.33, 95 % CI: 0.09-1.18) compared to men (HR = 2.09, 95 % CI: 0.78-5.61), p = 0.02 for interaction. CONCLUSIONS: In this large, pooled analysis of asymptomatic patients comparing CAS to CEA, sex did not act as an effect modifier of treatment differences in the four-year primary stroke-MI-or-death endpoint or the secondary stroke-or-death endpoint. However, during the post-procedural period men treated with CAS were at higher risk than their female counterparts.
Assuntos
Doenças Assintomáticas , Estenose das Carótidas , Endarterectomia das Carótidas , Índice de Gravidade de Doença , Stents , Acidente Vascular Cerebral , Humanos , Estenose das Carótidas/mortalidade , Estenose das Carótidas/terapia , Estenose das Carótidas/diagnóstico por imagem , Estenose das Carótidas/complicações , Estenose das Carótidas/cirurgia , Feminino , Masculino , Endarterectomia das Carótidas/efeitos adversos , Endarterectomia das Carótidas/mortalidade , Idoso , Fatores Sexuais , Resultado do Tratamento , Fatores de Risco , Acidente Vascular Cerebral/mortalidade , Acidente Vascular Cerebral/etiologia , Acidente Vascular Cerebral/diagnóstico , Fatores de Tempo , Medição de Risco , Pessoa de Meia-Idade , Ensaios Clínicos Controlados Aleatórios como Assunto , Infarto do Miocárdio/mortalidade , Infarto do Miocárdio/etiologia , Disparidades nos Níveis de Saúde , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/mortalidade , Procedimentos Endovasculares/instrumentaçãoRESUMO
OBJECTIVE: The objective of this study was to compare the rate of development of buttock claudication in patients undergoing aortoiliac aneurysm repair with and without exclusion of antegrade hypogastric arterial flow. In the absence of convincing data, questions remain regarding the best management of hypogastric arterial flow to prevent the theoretical risk of buttock claudication. METHODS: The Veterans' Affairs Open Versus Endovascular Repair (OVER) Cooperative Study prospectively collected information on buttock claudication. Trial participants were specifically prompted both pre- and postoperatively to report the development of claudication symptoms at several anatomic levels. Of note, trial investigators were specifically trained to occlude the trunk hypogastric arterial, preserving the anterior and posterior divisions. Bayesian survival models were created to evaluate time to development of left, right, or bilateral buttock claudication according to the presence/absence of antegrade hypogastric perfusion. RESULTS: A total of 881 patients from the OVER trial with information regarding status of hypogastric flow were included in the analysis. Of these, 788 patients maintained bilateral antegrade hypogastric arterial perfusion, 63 had right hypogastric coverage/occlusion, and 27 had left hypogastric coverage/occlusion, whereas 3 patients had bilateral hypogastric coverage/occlusion. Just under 5% of all patients (n = 41) developed buttock claudication. After adjustment for smoking, chronic obstructive pulmonary disease, medications, study arm, preoperative activity level, body mass index, age, and diabetes, intervention-related changes to hypogastric perfusion had no effect on time to development of buttock claudication. A Maximum A Posteriori Kullback- Leibler misfit χ2 was 14.45 with 24 degrees of freedom, resulting in a goodness of fit P-value of P = .94, indicative of a good fit. CONCLUSIONS: OVER is the largest aneurysm treatment study to prospectively collect data related to the development of claudication as well as hypogastric preservation status. Despite this, we were unable to find evidence to support the assertion that preservation of antegrade hypogastric flow decreases the rate of development of buttock claudication symptoms. The low rate of development of buttock claudication overall and in the subgroups is striking.
Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Ilíaco , Humanos , Aorta/cirurgia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/cirurgia , Teorema de Bayes , Implante de Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/métodos , Embolização Terapêutica/métodos , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/métodos , Aneurisma Ilíaco/diagnóstico por imagem , Aneurisma Ilíaco/cirurgia , Resultado do TratamentoRESUMO
OBJECTIVE: We have reported the 5-year results of a pivotal prospective, multicenter study conducted in the United States of a specifically designed iliac branch endoprosthesis (IBE; W.L. Gore & Associates, Flagstaff, AZ) for endovascular repair of aortoiliac aneurysms and common iliac artery aneurysms. METHODS: A total of 63 patients (98.4% male; mean age, 70 years) with aortoiliac or common iliac artery aneurysms had undergone implantation of a single IBE device and a bifurcated aortoiliac stent graft. Patients with bilateral common iliac artery aneurysms (n = 22; 34.9%) had undergone either staged occlusion or surgical revascularization of the contralateral internal iliac artery before study enrollment. At 5 years, 36 of the 63 patients had completed the final study follow-up examinations, including clinical examinations (n = 35) and computed tomography (n = 32), with the results evaluated by an independent core laboratory and adverse events adjudicated by a clinical events committee. RESULTS: At 5 years, freedom from all-cause mortality was 85.7% and freedom from aneurysm-related mortality was 100%. The nine deaths that had occurred (range, 132-1898 days) were adjudicated as unrelated to the aneurysm or procedure. Primary patency of the internal and external iliac artery IBE limbs was 95.1% and 100%, respectively. No patients had experienced new-onset buttock claudication on the IBE side or self-reported new-onset erectile dysfunction. The common iliac artery diameter on the IBE side was either unchanged or had decreased by ≥5 mm in 30 of the 31 patients (96.8%) with a baseline (1 month) and 5-year (range, 1641-2006 days) computed tomography scan available. Of the 31 evaluable patients, 9 (29.0%) had had an increase of ≥5 mm in the aortic diameter, 5 of whom had had a concurrent type II endoleak. No type I or type III endoleaks or device migration were identified by the core laboratory. Six patients had undergone eight secondary interventions, including five interventions for a type II endoleak. The freedom from secondary intervention was 90.5%. CONCLUSIONS: The 5-year results of our prospective, multicenter study have confirmed the safety, efficacy, and durability of the IBE device for the treatment of aortoiliac and iliac artery aneurysms. The device effectively prevented common iliac artery aneurysm rupture, maintained the patency of the internal iliac artery, and avoided the complications associated with internal iliac artery sacrifice. Although common iliac artery aneurysm enlargement was rare, abdominal aortic enlargement was more common, suggesting that the outcomes of endovascular aneurysm repair might be different for patients with or without associated common iliac artery aneurysms.
Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Procedimentos Endovasculares , Aneurisma Ilíaco , Humanos , Masculino , Idoso , Feminino , Prótese Vascular/efeitos adversos , Artéria Ilíaca/cirurgia , Endoleak/etiologia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/complicações , Estudos Prospectivos , Resultado do Tratamento , Aneurisma Ilíaco/diagnóstico por imagem , Aneurisma Ilíaco/cirurgia , Stents/efeitos adversos , Desenho de PróteseRESUMO
OBJECTIVE: Previous studies imply a profound residual mortality risk following successful abdominal aorta aneurysm (AAA) repair. This excess mortality is generally attributed to increased cardiovascular risk. The aim of this study was (1) to quantify the excess residual mortality for patients with AAA, (2) to evaluate the cross sectional level of cardiovascular risk management, and (3) to estimate the potential of optimised cardiovascular risk management to reduce the excess mortality in these patients. METHODS: Excess mortality was estimated through a systematic review and meta-analysis, and through data from the Swedish National Health Registry. Cardiovascular risk profiles were individually assessed during eligibility screening of patients with AAA for two multicentre pharmaceutical AAA stabilisation trials. The potential of full implementation of cardiovascular risk management was estimated through the validated Second Manifestations of ARTerial disease (SMART) risk scores algorithm. RESULTS: The meta-analysis showed a similarly impaired survival for patients who received early repair (small AAA) or regular repair (≥ 55 mm), and a further impaired survival for patients under surveillance for a small AAA. Excess mortality was further quantified using Swedish population data. The data revealed a more than quadrupled and doubled five year mortality rate for women and men who had their AAA repaired, respectively. Evaluation of the level of risk management of 358 patients under surveillance in 16 Dutch hospitals showed that the majority of patients with AAA did not meet therapeutic targets set for risk management in high risk populations, and indicated a more pronounced prevention gap in women. Application of the SMART risk score algorithm predicted that optimal implementation of risk management guidelines would reduce the 10 year risk of major adverse cardiovascular events from 43% to 14%. CONCLUSION: Independent of the rupture risk, AAA is associated with a worryingly compromised life expectancy with a particularly poor prognosis for women. Optimal implementation of cardiovascular risk prevention guidelines is predicted to profoundly reduce cardiovascular risk.
Assuntos
Aneurisma da Aorta Abdominal , Doenças Cardiovasculares , Masculino , Humanos , Feminino , Fatores de Risco , Estudos Transversais , Fatores de Risco de Doenças Cardíacas , Aneurisma da Aorta Abdominal/cirurgiaRESUMO
BACKGROUND: Published reports suggest that exclusion of antegrade hypogastric artery flow may have deleterious effects on erectile function after abdominal aortic aneurysm (AAA) repair. Off-label and open surgical hybrid procedures and, more recently, purpose-built branched devices have been developed to maintain antegrade pelvic perfusion in patients undergoing endovascular repair. Maintaining antegrade perfusion may reduce a spectrum of risks, including buttock claudication, colorectal ischemia, and spinal cord ischemia when patients undergo subsequent thoracic aortic procedures, as well as erectile dysfunction (ED). This project specifically focuses on erectile function, and analyzes baseline associations and relationships of hypogastric artery exclusion on changes in erectile function following aneurysm repair. METHODS: Male patients in the Veterans Affairs Open Versus Endovascular Repair (CSP#498; OVER) Trial had erectile function assessed preoperatively and postoperatively by administration of the International Index of Erectile Function-5 questionnaire. Bayesian mixed-effects regression models were created with the outcome variable (erectile function) treated as a latent variable. Primary effects of differences in erectile function between groups with and without preservation of bilateral antegrade hypogastric flow were compared. RESULTS: 876 men (442 randomized to endovascular repair) were enrolled in the trial and included in the analysis comparing treatment assignment. There is significant ED in elderly men with aortic aneurysm at baseline. Over 5 years of follow-up, there is modest decrease in erectile function and the endovascular group has improved function compared to open repair (0.082; 95% credible interval (CI) 0.008 and 0.155). A fifth of patients did not have bilateral preservation of antegrade hypogastric artery perfusion, with no difference in erectile function by univariate analysis. A more detailed regression analysis was applied--and after adjustment for baseline score, age, beta blocker use, diabetes, activity level, ejection fraction, preoperative ankle-brachial indices and time--preservation of both antegrade hypogastric arteries' perfusion showed transient improvement in survey scores compared to occlusion of at least 1 hypogastric artery at 6 months and 12 months after treatment, although this was not sustained at 60 months (score change: 0.046; 95% CI: -0.123, 0.215). Retesting this model in the cohort with complete data as a sensitivity analysis did not meaningfully change the conclusions. CONCLUSIONS: In this large prospective aneurysm treatment trial with systematic measurement of erectile function with a validated instrument, endovascular repair is associated with improved erectile function. Preservation of antegrade hypogastric flow with any repair is associated with early improved erectile function; however, it is not a sustained benefit. There is limited benefit of maintaining bilateral hypogastric artery perfusion for this specific indication in unselected men undergoing AAA repair.
Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Procedimentos Endovasculares , Disfunção Erétil , Aneurisma Ilíaco , Humanos , Masculino , Idoso , Lactente , Disfunção Erétil/etiologia , Disfunção Erétil/prevenção & controle , Estudos Prospectivos , Teorema de Bayes , Resultado do Tratamento , Artéria Ilíaca/cirurgia , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/cirurgia , Perfusão , Aneurisma Ilíaco/cirurgiaRESUMO
BACKGROUND: Doxycycline has been shown to prevent arterial calcification via attenuation of matrix metalloproteinases (MMP) in preclinical models. We assessed the effects of doxycycline on progression of arterial calcification in patients enrolled in the Non-Invasive Treatment of Abdominal Aortic Aneurysm Clinical Trial (N-TA3CT). METHODS: Two hundred and sixty-one patients were randomized to 100 mg doxycycline twice daily or placebo. Arterial calcification was measured in abdominal vessels on noncontrast computed tomography scans. Patients with baseline computed tomography scan and 1 or more follow-up scans within the 2-year study were included for analysis. For individual arteries, mean change in iliofemoral artery calcification over time was calculated via linear regression. Serum MMP-3 and MMP-9 levels were measured at baseline and 6 months. RESULTS: Sixty-five patients in the doxycycline and 66 in the placebo arm were included in this analysis. Baseline characteristics between the groups were similar. The unadjusted mean change in iliofemoral calcium score per year trended toward higher values in patients treated with doxycycline compared with placebo (322 ± 399 units/year vs. 217 ± 307 units/year, P = 0.09). After 6 months, changes in serum MMP-3 and MMP-9 levels were not significantly different between study arms. CONCLUSIONS: In patients with small aortic aneurysm, treatment with doxycycline 100 mg twice daily did not decrease circulating levels of the matrix degrading enzymes MMP-3 and 9 or alter the progression of arterial calcification.
RESUMO
BACKGROUND: Elective endovascular repair of an abdominal aortic aneurysm results in lower perioperative mortality than traditional open repair, but after 4 years this survival advantage is not seen; in addition, results of two European trials have shown worse long-term outcomes with endovascular repair than with open repair. Long-term results of a study we conducted more than a decade ago to compare endovascular repair with open repair are unknown. METHODS: We randomly assigned patients with asymptomatic abdominal aortic aneurysms to either endovascular repair or open repair of the aneurysm. All the patients were candidates for either procedure. Patients were followed for up to 14 years. RESULTS: A total of 881 patients underwent randomization: 444 were assigned to endovascular repair and 437 to open repair. The primary outcome was all-cause mortality. A total of 302 patients (68.0%) in the endovascular-repair group and 306 (70.0%) in the open-repair group died (hazard ratio, 0.96; 95% confidence interval [CI], 0.82 to 1.13). During the first 4 years of follow-up, overall survival appeared to be higher with endovascular repair than with open repair; from year 4 through year 8, overall survival was higher in the open-repair group; and after 8 years, overall survival was once again higher in the endovascular-repair group (hazard ratio for death, 0.94; 95% CI, 0.74 to 1.18). None of these trends were significant. There were 12 aneurysm-related deaths (2.7%) in the endovascular-repair group and 16 (3.7%) in the open-repair group (between-group difference, -1.0 percentage point; 95% CI, -3.3 to 1.4); most deaths occurred during the perioperative period. Aneurysm rupture occurred in 7 patients (1.6%) in the endovascular-repair group, and rupture of a thoracic aneurysm occurred in 1 patient (0.2%) in the open-repair group (between-group difference, 1.3 percentage points; 95% CI, 0.1 to 2.6). Death from chronic obstructive lung disease was just over 50% more common with open repair (5.4% of patients in the endovascular-repair group and 8.2% in the open-repair group died from chronic obstructive lung disease; between-group difference, -2.8 percentage points; 95% CI, -6.2 to 0.5). More patients in the endovascular-repair group underwent secondary procedures. CONCLUSIONS: Long-term overall survival was similar among patients who underwent endovascular repair and those who underwent open repair. A difference between groups was noted in the number of patients who underwent secondary therapeutic procedures. Our results were not consistent with the findings of worse performance of endovascular repair with respect to long-term survival that was seen in the two European trials. (Funded by the Department of Veteran Affairs Office of Research and Development; OVER ClinicalTrials.gov number, NCT00094575.).
Assuntos
Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/cirurgia , Procedimentos Endovasculares , Idoso , Aneurisma da Aorta Abdominal/mortalidade , Causas de Morte , Procedimentos Cirúrgicos Eletivos/métodos , Procedimentos Endovasculares/métodos , Seguimentos , Humanos , Estimativa de Kaplan-Meier , Complicações Pós-Operatórias , Resultado do TratamentoRESUMO
OBJECTIVE: Current management of small abdominal aortic aneurysms (AAAs) primarily involves serial imaging surveillance of maximum transverse diameter (MTD) to estimate rupture risk. Other measurements, such as volume and tortuosity, are less well-studied and may help characterize and predict AAA progression. This study evaluated predictors of AAA volume growth and discusses the role of volume in clinical practice. METHODS: Subjects from the Non-invasive Treatment of Abdominal Aortic Aneurysm Clinical Trial (baseline AAA MTD, 3.5-5.0 cm) with ≥2 computed tomography scans were included in this study (n = 250). Computed tomography scans were conducted approximately every 6 months over 2 years. MTD, volume, and tortuosity were used to model growth. Univariable and multivariable backwards elimination least squares regressions assessed associations with volume growth. RESULTS: Baseline MTD accounted for 43% of baseline volume variance (P < .0001). Mean volume growth rate was 10.4 cm3/year (standard deviation, 8.8 cm3/year) (mean volume change +10.4%). Baseline volume accounted for 30% of volume growth variance; MTD accounted for 13% of volume growth variance. More tortuous aneurysms at baseline had significantly larger volume growth rates (difference, 32.8 cm3/year; P < .0001). Univariable analysis identified angiotensin II receptor blocker use (difference, -3.4 cm3/year; P = .02) and history of diabetes mellitus (difference, -2.8 cm3/year; P = .04) to be associated with lower rates of volume growth. Baseline volume, tortuosity index, current tobacco use, and absence of diabetes mellitus remained significantly associated with volume growth in multivariable analysis. AAAs that reached the MTD threshold for repair had a wide range of volumes: 102 cm3 to 142 cm3 in female patients (n = 5) and 105 cm3 to 229 cm3 in male patients (n = 20). CONCLUSIONS: Baseline AAA volume and MTD were found to be moderately correlated. On average, AAA volume grows about 10% annually. Baseline volume, tortuosity, MTD, current tobacco use, angiotensin II receptor blocker use, and history of diabetes mellitus were predictive of volume growth over time.
Assuntos
Aneurisma da Aorta Abdominal , Antagonistas de Receptores de Angiotensina , Aneurisma da Aorta Abdominal/complicações , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Feminino , Humanos , Masculino , Tomografia Computadorizada por Raios XRESUMO
OBJECTIVE: Asymptomatic carotid stenosis is the most frequent indication for carotid endarterectomy (CEA) in the United States. Published trials and guidelines support CEA indications in selected patients with longer projected survival and when periprocedural complications are low. Transfemoral carotid artery stenting with embolic protection (CAS) is a newer treatment option. The objective of this study was to compare outcomes in asymptomatic, nonoctogenarian patients treated with CAS vs CEA. METHODS: Patient-level data was analyzed from 2544 subjects with ≥70% asymptomatic carotid stenosis who were randomized to CAS or CEA in addition to standard medical therapy. One trial enrolled 1091 (548 CAS, 543 CEA) and another enrolled 1453 (1089 CAS, 364 CEA) asymptomatic patients less than 80 years old (upper age eligibility). Independent neurologic assessment and routine cardiac enzyme screening were performed. The prespecified, primary composite endpoint was any stroke, myocardial infarction, or death during the periprocedural period or ipsilateral stroke within 4 years after randomization. RESULTS: There was no significant difference in the primary endpoint between CAS and CEA (5.3% vs 5.1%; hazard ratio, 1.02; 95% confidence interval, 0.7-1.5; P = .91). Periprocedural rates for the components are (CAS vs CEA): any stroke (2.7% vs 1.5%; P = .07), myocardial infarction (0.6% vs 1.7%; P = .01), death (0.1% vs 0.2%; P = .62), and any stroke or death (2.7% vs 1.6%; P = .07). After this period, the rates of ipsilateral stroke were similar (2.3% vs 2.2%; P = .97). CONCLUSIONS: In a pooled analysis of two large randomized trials of CAS and CEA in asymptomatic, nonoctogenarian patients, CAS achieves comparable short- and long-term results to CEA.
Assuntos
Estenose das Carótidas , Endarterectomia das Carótidas , Infarto do Miocárdio , Acidente Vascular Cerebral , Idoso de 80 Anos ou mais , Humanos , Estenose das Carótidas/complicações , Estenose das Carótidas/diagnóstico por imagem , Estenose das Carótidas/cirurgia , Endarterectomia das Carótidas/efeitos adversos , Infarto do Miocárdio/etiologia , Medição de Risco , Fatores de Risco , Stents/efeitos adversos , Acidente Vascular Cerebral/etiologia , Acidente Vascular Cerebral/prevenção & controle , Resultado do Tratamento , Estados UnidosRESUMO
OBJECTIVE: Thoracic endovascular aortic repair has emerged as the dominant paradigm for treatment of patients with descending thoracic aortic aneurysms. For aneurysms involving the aortic arch in the region of the left subclavian artery (LSA), branch vessel preservation to maintain blood flow to the LSA is recommended. Branched aortic endografts are an alternative to surgical revascularization of the LSA. METHODS: Across 34 investigative sites, 84 patients with zone 2 aneurysms were enrolled in a nonrandomized, prospective study of a single branched aortic endograft. The thoracic branch endoprosthesis device allows for graft placement proximal to the LSA and incorporates a single side branch for left subclavian perfusion. RESULTS: More than one-half of the patients were male (63%). Their average age was 70 ± 11 years. The aneurysm morphology was fusiform in 43 and saccular in 41 patients. The mean aneurysm diameter at screening was 56.2 mm. The mean follow-up was 30 months (range, 2.6-50.7 months). Reported here are the patient outcomes at 1 and 12 months. Predefined technical success with implantation of the device in landing zone 2 was achieved in 92% of patients (n = 77). There were no cases of aortic rupture, lesion-related mortality, or new-onset renal failure. There was no perioperative (30-day) mortality. A single case each of permanent paraplegia and paraparesis occurred. Three patients experienced a procedure-related stroke. Through 12 months, four patients died; none of the deaths were adjudicated as related to the device or procedure. One aortic reintervention was required. A single case of aortic enlargement (core laboratory) was reported at 6 months. Type I (n = 3) and III (n = 5) endoleaks occurred in 9.8% of patients, of which one (type III) required reintervention. CONCLUSIONS: Results from this device study in patients with zone 2 aneurysms demonstrate that early safety and efficacy outcomes are maintained up to 12 months after the endovascular procedure with low mortality and reintervention rates and an acceptable frequency of procedural complications, including neurologic complications.
Assuntos
Aneurisma da Aorta Torácica , Ruptura Aórtica , Implante de Prótese Vascular , Procedimentos Endovasculares , Humanos , Masculino , Pessoa de Meia-Idade , Idoso , Idoso de 80 Anos ou mais , Feminino , Artéria Subclávia/diagnóstico por imagem , Artéria Subclávia/cirurgia , Aneurisma da Aorta Torácica/diagnóstico por imagem , Aneurisma da Aorta Torácica/cirurgia , Aneurisma da Aorta Torácica/etiologia , Prótese Vascular , Estudos Prospectivos , Resultado do Tratamento , Aorta Torácica/cirurgia , Ruptura Aórtica/cirurgia , Estudos RetrospectivosRESUMO
OBJECTIVE: Aortic aneurysms involving aortic arch vessels are anatomically unsuitable for standard thoracic endovascular repair (TEVAR) without cervical debranching of the arch vessels. Three year outcomes of a single branched thoracic endograft following previous publication of peri-operative and one year outcomes are reported. METHODS: This was a multicentre feasibility trial of the GORE TAG Thoracic Branch Endoprosthesis (TBE), a thoracic endovascular graft incorporating a single retrograde branch for aortic arch vessel perfusion. The first study arm enrolled patients with an intact descending thoracic aortic aneurysm extending to the distal arch with left subclavian artery (LSA) incorporation (zone 2). The second arm enrolled patients with arch aneurysms requiring incorporation of the left carotid or innominate artery (zone 0/1) and extra-anatomic surgical revascularisation of the remaining aortic arch vessels. Outcomes at three years are reported. RESULTS: The cohort comprised 40 patients (31 zone 2, nine zone 0/1). The majority were male (52%). Mean follow up was 1 408 ± 552 days in the zone 2 and 1 187 ± 766 days in the zone 0/1 cohort. During three year follow up there was no device migration, fracture, or aortic rupture in either arm. In the zone 2 arm, freedom from re-intervention was 97% at one and three years but there were two side branch occlusions. Two patients had aneurysm enlargement > 5 mm without documented endoleak or re-intervention. Freedom from death at one and three years was 90% and 84%. In the zone 0/1 arm there were no re-interventions, loss of branch patency, or aneurysm enlargement at three years. Cerebrovascular events occurred in three patients during follow up: two unrelated to the device or procedure, and one of unknown relationship. Two patients in this arm died during the follow up period, both unrelated to the procedure or the aneurysm. CONCLUSION: Initial three year results of the TBE device for endovascular repair of arch aneurysms show favourable patency and durability with low rates of graft related complications.
Assuntos
Aneurisma do Arco Aórtico , Aneurisma da Aorta Torácica , Aneurisma Aórtico , Implante de Prótese Vascular , Procedimentos Endovasculares , Feminino , Humanos , Masculino , Aorta Torácica/diagnóstico por imagem , Aorta Torácica/cirurgia , Aneurisma Aórtico/cirurgia , Aneurisma da Aorta Torácica/diagnóstico por imagem , Aneurisma da Aorta Torácica/cirurgia , Prótese Vascular , Complicações Pós-Operatórias/cirurgia , Desenho de Prótese , Estudos Retrospectivos , Stents , Resultado do Tratamento , Estudos de ViabilidadeRESUMO
BACKGROUND: The Gore conformable TAG thoracic endoprosthesis (CTAG) was engineered to enhance conformability in the thoracic aorta and has demonstrated greater deployment accuracy and wall apposition than the original TAG device. The Food and Drug Administration approved the CTAG in 2011 based on the 1-year results of the pivotal trial for the CTAG device. In the present report, we have documented the outcomes for those patients through 5 years. METHODS: The CTAG aneurysm study was a prospective, multicenter trial that assessed the safety and effectiveness of the CTAG device as treatment of descending thoracic aortic aneurysms. Follow-up imaging assessments consisted of radiographs and computed tomography at annual intervals through 5 years. A core imaging laboratory was used to assess the aneurysm size, device integrity, and endoleaks. RESULTS: A total of 66 patients were enrolled (51 pivotal arm patients and 15 continued access patients) from October 2009 through September 2011. The baseline characteristics, procedural characteristics, and outcomes through 2 years were reported previously. Through 5 years, five patients (7.6%) had required device-related reintervention (one type Ia endoleak or contained rupture of the descending thoracic aorta, one type Ib endoleak, two indeterminate endoleaks, and one thoracic aortic pseudoaneurysm). Four patients (6.1%) had died of aneurysm-related causes, with one death occurring within 30 days of the index procedure. A total of 24 patients (36.3%) had died during the 5-year study period. Seven patients (10.6%) had experienced stroke or transient ischemic attack through 5 years, with one early stroke (postoperative day 28). Prosthesis or intercomponent migration was observed in six patients (9.1%) through 5 years; however, no patient had developed a type III junctional endoleak. No stent-graft fracture or compression was observed through 5 years. At 5 years, 14 patients (50.0%) showed sac regression, 5 (17.9%) showed sac expansion, and 9 (32.1%) showed sac stability using on a 5-mm threshold for change. CONCLUSIONS: Thoracic endovascular aortic repair with the CTAG device was associated with low rates of aneurysm-related mortality and reintervention through 5 years. Proximal endoleak was rare, and most patients showed sac regression or stability at 5 years after the initial thoracic endovascular aortic repair.