Pediatr Emerg Care ; 38(2): e731-e733, 2022 Feb 01.
Artigo em Inglês | MEDLINE | ID: mdl-33394949

RESUMO

OBJECTIVE: To determine the motions produced during pediatric extrication when using a system of motion estriction and extrication. METHODS: Simulation-based biomechanical analysis study conducted with inertial sensors to measure motion produced in the cervical spine of a pediatric simulator during extrication from a vehicle. RESULTS: The mean of the movements was 3.5° (SD ±1.35°). The mean time was 4 minutes 1 second (SD, ±45.09 seconds). The mean rotation toward the right was 3.34° (SD ±3.52°) and toward the left 2.62° (SD ±2.26°). The mean for lateralization was 6.24° (SD ±3.20°) toward the right and 2.50° (DE ±2.76°) toward the left. The mean for flexion was 2.36° (SD ±2.10°) and for extension 4.21° (SD ±2.15°). CONCLUSIONS: The device analyzed allows for the extrication of the pediatric patient with high levels of motion restriction of the spinal column with the Pediatric Immobilization and Extrication System.


Assuntos
Vértebras Cervicais , Imobilização , Criança , Humanos , Movimento , Amplitude de Movimento Articular , Rotação