Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Mais filtros

Base de dados
País/Região como assunto
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
J Endovasc Ther ; : 15266028241248311, 2024 Apr 23.
Artigo em Inglês | MEDLINE | ID: mdl-38654524

RESUMO

PURPOSE: Recent reports of the Zenith Alpha abdominal endovascular graft (Zenith Alpha AAA, Cook, Inc., Bloomington, Indiana) have demonstrated an unexpectedly high incidence of limb graft occlusion (LGO). The purpose of this study was to prospectively evaluate the performance of the Zenith AAA in Japan, with a specific focus on LGO. MATERIALS AND METHODS: All endovascular aneurysm repairs (EVARs) for abdominal aortic aneurysms performed using the Zenith Alpha AAA from July 2020 to October 2021 in 23 Japanese hospitals were prospectively evaluated. All computed tomographic images were analyzed in the core laboratory. Late complications were defined as any aneurysm-related events occurring >30 days after EVAR, including aneurysm sac enlargement of >5 mm and any reinterventions performed. Endoleaks without sac enlargement or reintervention were excluded as late complications. RESULTS: During the study period, 147 EVARs were performed using the Zenith Alpha AAA. The mean patient age was 76.5±7.7 years, 84.4% of patients were male, and the mean aneurysm diameter was 52.4±9.2 mm. Instruction-for-use violations were observed in 76 patients (51.7%), primarily associated with a severely angulated proximal neck (>60°). There were six intraoperative complications and 62 additional intraoperative treatments reported, most of which involved preemptive coil embolization of the inferior mesenteric artery (37 cases). Technical success, defined as the absence of type 1 or 3 endoleaks on final angiography, was achieved in 99.3% of patients. At 12 months, there was only one case of type 1/3 endoleak (0.8%) and one aneurysm sac enlargement exceeding 5 mm (0.8%); however, a high incidence of type 2 endoleaks was observed in 35.2% of patients, and aneurysm sac regression exceeding 5 mm was achieved in 30.1% of patients. Nine late complications were observed, and the rate of freedom from late complications at 12 months was 93.5%, encompassing four LGOs and one limb graft stenosis (3.4%). CONCLUSIONS: In contrast to recent reports, our Japanese multicenter prospective study demonstrated satisfactory early clinical results, including an acceptable LGO rate, for the low-profile Zenith Alpha AAA. Long-term follow-ups will be performed to confirm the persistence of these outcomes. CLINICAL IMPACT: This study prospectively evaluated the performance of 147 Zenith Alpha AAAs used for endovascular aneurysm repair with core-lab adjudication focusing especially on limb graft occlusion (LGO). At 12 months, aneurysm sac regression exceeding 5 mm was achieved in 30.1% of patients, and there was only one type 1/3 endoleak, one aneurysm sac enlargement (>5 mm), and nine late complications including five LGOs (3.4%), resulting in rate of freedom from late complications at 93.5%. Satisfactory early clinical results, including an acceptable rate of LGO can be achieved, particularly with cautious usage of the Zenith Alpha Spiral-Z Endovascular Leg." 5mm), and nine late complications including five LGOs (3.4%), resulting in rate of freedom from late complications at 93.5%. Satisfactory early clinical results, including an acceptable rate of LGO can be achieved, particularly with cautious usage of the Zenith Alpha Spiral-Z Endovascular Leg.

2.
J Vasc Surg ; 77(3): 770-777.e2, 2023 03.
Artigo em Inglês | MEDLINE | ID: mdl-36306934

RESUMO

OBJECTIVE: Prior reports of the low profile Zenith Alpha abdominal graft (Cook Medical Inc, Bloomington, IN) have shown impaired limb graft patency to be the primary causes of reintervention. Special notices from the manufacturer have indicated certain instructions for use (IFU) violations as the main reasons for these complications. In the present study, we assessed the incidence of limb graft occlusion (LGO) and analyzed the effects of the detailed anatomic risk factors for LGO highlighted in the IFU and previously reported studies. METHODS: A retrospective study was performed of 241 patients treated with the low profile Zenith Alpha at a single institution from October 1, 2015 to September 30, 2018. All computed tomography angiograms were analyzed using three-dimensional software. Data were extracted from the electronic medical records until the end of the study period (December 31, 2020). The cumulative incidence of LGO and LGO-related reinterventions were assessed. A regression analysis was performed to evaluate the possible risk factors associated with the development of LGO at specified time points. These included aortic and iliac diameters, graft component oversizing, iliac tortuosity and calcification, overlap of graft components, proximal alignment of ipsilateral and contralateral legs, and sealing zone in the external iliac artery. Reader agreement of iliac calcification and tortuosity was assessed in patients with LGO. RESULTS: A total of 33 limbs (7%) in 27 patients (11%) had become occluded. The cumulative incidence of LGO was 7% (95% confidence interval [CI], 5%-9%) per limb up to 3 years postoperatively. The previously described risk factors for LGO were studied using regression analysis; however, no positive association with LGO was identified. Heavily calcified common iliac arteries (CIAs) and external iliac arteries were protective against LGO compared with noncalcified vessels up to 3 years postoperatively (decreased risk, 17% [95% CI, -27% to -7%]; P = .001; and 15% [95% CI, -26 to -5]; P = .005, respectively). The reader agreement of iliac calcification and tortuosity showed substantial agreement (CIA intrareader kappa = 0.75; CIA interreader kappa = 0.62) and almost perfect agreement (intrareader kappa = 0.85; interreader kappa = 0.84), respectively. CONCLUSIONS: The cumulative incidence of LGO after endovascular aneurysm repair with the Zenith Alpha graft was 7% per limb up to 3 years postoperatively. None of the analyzed risk factors suggested by the IFUs or current literature were positively associated with LGO.


Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Procedimentos Endovasculares , Humanos , Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/efeitos adversos , Correção Endovascular de Aneurisma , Stents , Aneurisma da Aorta Abdominal/cirurgia , Estudos Retrospectivos , Resultado do Tratamento , Procedimentos Endovasculares/efeitos adversos , Fatores de Risco , Desenho de Prótese
3.
J Endovasc Ther ; 30(2): 241-249, 2023 04.
Artigo em Inglês | MEDLINE | ID: mdl-35249411

RESUMO

PURPOSE: This study retrospectively evaluated the 12-month outcomes of the Zenith Alpha Abdominal Endovascular graft (Zenith Alpha AAA, Cook, Inc, Bloomington, Indiana), which was launched partially in Japan in March 2019, starting with 9 selected sites. METHODS: A retrospective analysis was performed of all endovascular aneurysm repairs (EVAR) for abdominal aortic aneurysms using the Zenith Alpha AAA. Late complications were defined as any aneurysm-related events occurring >30 days after EVAR, including an aneurysm sac enlargement of >5 mm and any reinterventions performed. Endoleaks without sac enlargement or reinterventions were excluded from late complications. RESULTS: During the study period, 79 EVARs using the Zenith Alpha AAA were performed. The mean age was 76.6±6.9 years old, 91.1% of patients were male, and the mean aneurysm diameter was 51.1±7.5 mm. Instructions for use violation was observed in 27 patients (34.2%), most frequently being a severely angulated proximal neck (>60°). There were 4 intraoperative complications, including 2 unintentional partial renal artery coverages caused by the premature anchoring of the exposed suprarenal stent barb to the aortic wall, which was a result of the design change of the top cap deletion. Since it was a partial coverage without flow impairment and since renal stenting was unsuccessful, it was left untreated and had no subsequent renal function impairment. During the mean follow-up of 444±123 days, 74 patients completed 12 months of follow-up. Freedom from late complications at 12 months was 90.8%, which included 2 limb occlusions (2.5%). Of 71 patients with a 12-month computed tomography scan, there was only 1 type 1a endoleak (1.3%), 1 sac enlargement of >5 mm (1.3%), and an aneurysm sac shrinkage of >5 mm was observed in 42.2% of patients. There was no type 3 endoleak during the follow-up. CONCLUSIONS: This study demonstrated that the new generation of low-profile Zenith Alpha AAA has satisfactory early clinical outcomes, comparable to those obtained with the conventional Zenith endovascular graft. Long-term follow-up is needed to determine whether these favorable outcomes persist.


Assuntos
Aneurisma da Aorta Abdominal , Implante de Prótese Vascular , Procedimentos Endovasculares , Humanos , Masculino , Idoso , Idoso de 80 Anos ou mais , Feminino , Prótese Vascular/efeitos adversos , Estudos Retrospectivos , Implante de Prótese Vascular/efeitos adversos , Implante de Prótese Vascular/métodos , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/cirurgia , Aneurisma da Aorta Abdominal/complicações , Endoleak/diagnóstico por imagem , Endoleak/etiologia , Resultado do Tratamento , Japão , Fatores de Risco , Procedimentos Endovasculares/efeitos adversos , Procedimentos Endovasculares/métodos
4.
J Vasc Surg ; 76(4): 942-950.e1, 2022 10.
Artigo em Inglês | MEDLINE | ID: mdl-35367569

RESUMO

OBJECTIVE: Outcome reports after endovascular aneurysm repair (EVAR) using the low-profile Zenith Alpha Abdominal Endovascular grafts (Cook Medical, Bjæverskov, Denmark) are sparse. We present results from a single-center cohort treated with the Zenith Alpha, from a period where the graft was the primary EVAR device choice. The aim of the study was to evaluate short- and midterm outcomes of patients treated with the Zenith Alpha. METHODS: A retrospective single-center study was performed including all patients treated with the Zenith Alpha graft from October 1, 2015 to September 30, 2018. All patients underwent computed tomography angiography (CTA) imaging preoperatively as well as at 3 and 12 months postoperatively. Hereafter, patients were followed yearly with duplex ultrasound and clinical exams. Additional imaging was performed on indication. All CTAs were analyzed using three-dimensional reconstruction software (Aquarius, TeraRecon, Durham, NC). Data was extracted from electronic charts according to a protocol that remained unchanged until the end of the study (December 31, 2020). The following outcomes were assessed according to Society for Vascular Surgery/International Society of Cardiovascular Surgeons reporting criteria: aortic-related and all-cause mortality, reinterventions, instruction for use (IFU) violations, endoleaks (ELs), and aneurysm shrinkage. RESULTS: A total of 241 patients were treated with the Zenith Alpha, and 214 (89%) were asymptomatic repairs. Technical success was achieved in 238 patients (99%). One hundred fifty-seven patients (65%) received implantation outside IFU. The median hospital length of stay was 2 days (interquartile range, 2-3 days). The median clinical follow-up was 35.1 months (interquartile range, 28.8-47.5 months). The 4-year Kaplan-Meier estimate of freedom from reintervention was 66% (95% confidence interval, 59%-73%). The main reasons for reinterventions were iliac limb stenosis and occlusion (n = 30; 12%) and type 2 EL (n = 13; 5%). Overall, significantly more patients with grafts implanted outside distal IFU developed type 1B ELs (n = 10/11; P = .009). Aneurysm sac shrinkage was observed in 48 patients (25%) 1 year postoperatively. The Kaplan-Meier estimate of freedom from aortic-related mortality was 99% (95% confidence interval, 98%-100%) 4 years postoperatively. CONCLUSIONS: EVAR with the Zenith Alpha shows acceptable freedom from aortic-related mortality up to 4 years postoperatively. The majority of patients were treated outside IFU, and significantly more type 1B ELs appeared in this subgroup of patients. The leading cause for reintervention was impaired limb patency. The root cause for impaired limb patency requires further investigation.


Assuntos
Aneurisma da Aorta Abdominal , Ruptura Aórtica , Implante de Prótese Vascular , Procedimentos Endovasculares , Aneurisma da Aorta Abdominal/complicações , Aneurisma da Aorta Abdominal/diagnóstico por imagem , Aneurisma da Aorta Abdominal/cirurgia , Ruptura Aórtica/cirurgia , Prótese Vascular/efeitos adversos , Endoleak/diagnóstico por imagem , Endoleak/etiologia , Endoleak/cirurgia , Humanos , Desenho de Prótese , Estudos Retrospectivos , Fatores de Risco , Stents/efeitos adversos , Fatores de Tempo , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA