Systematic errors in estimating mean blood pressure from finger blood pressure measurements.
O'Callaghan, C J; Straznicky, N E; Komersova, K; Louis, W J.
Afiliação
  • O'Callaghan CJ; Department of Medicine, University of Melbourne, Austin and Repatriation Medical Centre, Heidelberg, Australia. cjo1@le.ac.uk
Blood Press ; 7(5-6): 277-81, 1998 Nov.
Article em En | MEDLINE | ID: mdl-10321439
The mean blood pressure (BP) can be accurately estimated from indirect measurements of brachial artery pressure, i.e. mean BP = diastolic BP + 1/3 pulse pressure. Although this equation has been used as a surrogate of mean systemic pressure, it is unknown whether this approximation can be validly applied to distal vascular beds. Therefore we determined the accuracy of this method as an estimate of the mean pressure in distal arteries by measuring finger BP with the Finapres device in 16 normotensive and 12 hypertensive subjects. The "calculated" and measured values of mean BP were compared when subjects were resting and during manoeuvres which aimed to alter the shape of the pulse waveform. Although closely correlated with the measured value, the "calculated" resting mean BP was systematically greater (+2.7+/-0.7 mm Hg, p<0.001). Additionally, the rise in the mean pressure produced by infusion of phenylephrine, an alpha1-adrenoceptor stimulant (16.0+/-1.5 mm Hg) was underestimated by the calculation (13.1+/-1.5 mm Hg, p<0.05). Of even greater concern was that calculating the mean pressure during infusion of isoprenaline (a beta-adrenergic stimulant) suggested the mean pressure had increased by 5.8+/-1.6 mm Hg when it had actually fallen (-2.1+/-2.4 mm Hg, p<0.001 vs. the measured value). Thus, calculating the mean BP from Finapres measurements roughly approximates the measured value when subjects are at rest. However, this estimation becomes inaccurate when pulse wave dimensions are altered, and is probably unsuitable for assessing the acute effects of vasoactive drugs, in particular vasodilators, on BP.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Dedos Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Assunto da revista: ANGIOLOGIA Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Austrália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Dedos Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Assunto da revista: ANGIOLOGIA Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Austrália