An isocentrically mounted stand for total body irradiation.
Br J Radiol
; 73(871): 776-9, 2000 Jul.
Article
em En
| MEDLINE
| ID: mdl-11089471
The purpose of this study was to construct a stand to support a patient for total body photon irradiation and to expedite the set-up and treatment by rotating the stand. As in other isocentric treatments, the midline dose is impacted less by source-to-skin distance variations. The method of immobilizing the patient is described. A 10 mm lucite plate is supported in front of the patient to increase skin dose. A matrix of holes in this plate serves to index the location of blocks used to shield the lungs. The dosimetry of the set-up is described, as is the production of tissue deficit compensators. The results of phantom studies and in vivo thermoluminescent dosimetry measurements are presented.