Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium.
Carrasco, P; Jornet, N; Duch, M A; Weber, L; Ginjaume, M; Eudaldo, T; Jurado, D; Ruiz, A; Ribas, M.
Afiliação
  • Carrasco P; Servei de Radiofisica i Radioprotecció, Hospital de la Santa Creu i Sant Pau, St Antoni Maria Claret-167, 08025 Barcelona, Spain.
Med Phys ; 31(10): 2899-911, 2004 Oct.
Article em En | MEDLINE | ID: mdl-15543799
ABSTRACT
An extensive set of benchmark measurement of PDDs and beam profiles was performed in a heterogeneous layer phantom, including a lung equivalent heterogeneity, by means of several detectors and compared against the predicted dose values by different calculation algorithms in two treatment planning systems. PDDs were measured with TLDs, plane parallel and cylindrical ionization chambers and beam profiles with films. Additionally, Monte Carlo simulations by means of the PENELOPE code were performed. Four different field sizes (10 x 10, 5 x 5, 2 x 2, and 1 x 1 cm2) and two lung equivalent materials (CIRS, p(w)e=0.195 and St. Bartholomew Hospital, London, p(w)e=0.244-0.322) were studied. The performance of four correction-based algorithms and one based on convolution-superposition was analyzed. The correction-based algorithms were the Batho, the Modified Batho, and the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system and the TMS Pencil Beam from the Helax-TMS (Nucletron) treatment planning system. The convolution-superposition algorithm was the Collapsed Cone implemented in the Helax-TMS. The only studied calculation methods that correlated successfully with the measured values with a 2% average inside all media were the Collapsed Cone and the Monte Carlo simulation. The biggest difference between the predicted and the delivered dose in the beam axis was found for the EqTAR algorithm inside the CIRS lung equivalent material in a 2 x 2 cm2 18 MV x-ray beam. In these conditions, average and maximum difference against the TLD measurements were 32% and 39%, respectively. In the water equivalent part of the phantom every algorithm correctly predicted the dose (within 2%) everywhere except very close to the interfaces where differences up to 24% were found for 2 x 2 cm2 18 MV photon beams. Consistent values were found between the reference detector (ionization chamber in water and TLD in lung) and Monte Carlo simulations, yielding minimal differences (0.4%+/-1.2%). The penumbra broadening effect in low density media was not predicted by any of the correction-based algorithms, and the only one that matched the experimental values and the Monte Carlo simulations within the estimated uncertainties was the Collapsed Cone Algorithm.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Software / Radioterapia Conformacional / Pulmão / Modelos Biológicos Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Espanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Software / Radioterapia Conformacional / Pulmão / Modelos Biológicos Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Espanha