A dynamic renal phantom for nuclear medicine studies.
SabbirAhmed, A S M; Demir, Mustafa; Kabasakal, Levent; Uslu, Ilhami.
Afiliação
  • SabbirAhmed AS; Nukleer Tip ABD, Istanbul Universitesi, Cerrahpasa Tip Fakultesi, 34303 Cerrahpasa, Istanbul, Turkey.
Med Phys ; 32(2): 530-8, 2005 Feb.
Article em En | MEDLINE | ID: mdl-15789599
ABSTRACT
Dynamic radionuclide renal study (renography) provides functional and structural information of the kidney and urinary tract noninvasively. Our purpose in this study is to describe the construction and test results of a dynamic renal phantom with different clinical features of radionuclide renography. The phantom consisted of five pieces of different shaped Plexiglas boxes Two kidneys, one liver, two square shaped boxes (one heart and one bladder). The bladder was internally divided into two compartments in order to collect each kidney output separately. The dynamic circulation of the phantom was maintained under a hydrostatic pressure approximately equal to 13.3 kPa (average human blood pressure). The standard dose distribution among different organs and different renographic parameters were calculated from series of normal patients study (91 with 99mTc-DTPA, 68 with 99mTc-EC). All the studies were performed with same camera (Siemens Orbiter Digitrac 7500) equipped with LEAP (low energy all purpose) collimator using ADAC Pegasys II analytic package program under the same clinical procedure. Different regions of interest (ROIs) were drawn for concerning organs and counts per second (CPS) were collected for each ROI. The series of renogram curves were generated by phantom-studies with different flow rates for left kidney (LK) and right kidney (RK). The renal index (RI) for an individual study was calculated as the product of two indexes "Relative Renal Function" (RRF) (water-volume of LK/RK) and "Relative Renal Time" (RRT) (Tmax of LK/RK). The most significant correlation was found in total CPS for LK and RK between the EC group and phantom studies (p < 0.001). The calculated RI values were used to simulate the patients' study with different clinical features. The dynamics were found reproducible. The phantom is suitable for using in calibration and quality control protocols of the renogram procedure used in Nuclear Medicine.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Renografia por Radioisótopo / Imagens de Fantasmas / Análise de Falha de Equipamento / Rim / Medicina Nuclear Tipo de estudo: Guideline Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Med Phys Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Turquia
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Renografia por Radioisótopo / Imagens de Fantasmas / Análise de Falha de Equipamento / Rim / Medicina Nuclear Tipo de estudo: Guideline Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Med Phys Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Turquia