Fabrication of a hydrogen peroxide biosensor based on a self-assemble composite oxide film.
Front Biosci
; 10: 2841-7, 2005 Sep 01.
Article
em En
| MEDLINE
| ID: mdl-15970538
A titania film with horseradish peroxidase (HRP) was immobilized on indium tin oxide (ITO) electrodes to act as a biosensor of hydrogen peroxide via the self-assembling and sol-gel route. The film was characterized by the atomic functional microscopy, electrochemical and UV-vis spectroscopy methods. The resulting biosensor exhibited fast amperometric response and good stability to hydrogen peroxide (H2O2). The linear range for H2O2 determination was from 2 x10(-6) to 1.45 x 10(-3) M, with a detection limit of 4 x10(-6) M based on S/N=3. The apparent Michaelis-Menten constant of the biosensor to H2O2 was estimated to be 1.5 mM, showing the HRP on the films had the higher biologic activities.