Comparison of an automated and point-of-care immunoassay to GC-MS for urine oxycodone testing in the clinical laboratory.
Haller, Christine A; Stone, Judith; Burke, Vicki; Branch, Jean; Chen, Kathy; Gross, Susan.
Afiliação
  • Haller CA; Department of Medicine and Laboratory Medicine, University of California-San Francisco, Box 1220, San Francisco, CA 94143, USA. dchaller@worldnet.att.net
J Anal Toxicol ; 30(2): 106-11, 2006 Mar.
Article em En | MEDLINE | ID: mdl-16620541
ABSTRACT
OxyContin, a controlled-release formulation of oxycodone, is increasingly abused. Monitoring patient compliance by urine drug testing may deter illegal diversion of OxyContin. Two urine immunoassays were evaluated with a 100 ng/mL cutoff for oxycodone. The Microgenics Corporation Oxycodone DRI on the Bayer ADVIA 1650 and a point-of-care (POC) immunoassay, Monitect Oxycodone POC from Branan Medical Corporation, were compared to gas chromatography-mass spectrometry (GC-MS) with a detection limit of 50 ng/mL free oxycodone. Between-day precision for DRI yielded coefficients of variation from 3.9% to 7.0% at 75 and 125 ng/mL. Fifty-two positive and 52 negative urines were tested. The DRI had a 100% agreement with GC-MS. Two positive specimens had free oxycodone < 50 ng/mL, but oxycodone metabolites, oxymorphone and oxycodone glucuronide > 100 ng/mL, were identified by GC-MS analysis. The POC assay had two false positives and 15 indeterminate (+/-) results. Codeine or hydrocodone was present in all but one of these samples. There was no interference with DRI from morphine, codeine, hydrocodone, hydromorphone, dihydrocodeine, or 6-monoacetyl morphine. Four-hundred and ninety urine samples were subsequently tested with DRI to estimate the oxycodone-positive rate at our hospital, and 47 (9.4%) were positive. The confirmation rate with GC-MS for free oxycodone, not including metabolites, was 93%. The Microgenics DRI offers good performance for oxycodone urine testing and is a better choice for the clinical laboratory than the POC assay. Confirmation of screened positive samples requires a method that can detect total oxycodone and oxymorphone.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxicodona / Imunoensaio / Detecção do Abuso de Substâncias / Técnicas de Laboratório Clínico / Cromatografia Gasosa-Espectrometria de Massas / Entorpecentes Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Anal Toxicol Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxicodona / Imunoensaio / Detecção do Abuso de Substâncias / Técnicas de Laboratório Clínico / Cromatografia Gasosa-Espectrometria de Massas / Entorpecentes Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Anal Toxicol Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Estados Unidos