Your browser doesn't support javascript.
loading
[Adaptation to the Biomek 1000 workstation (Beckman) of the MHA-TP technique (Ames)]. / L'adaptation à l'automate Biomek 1000 (Beckman) de la technique de TPHA (Ames).
Le Guyader, F; Segui, A; Minet, J; Guerin, J; Cormier, M.
Afiliação
  • Le Guyader F; Laboratoire de Bactériologie et Sérologie Bactériologienne, CHRU, Rennes.
Ann Biol Clin (Paris) ; 49(5): 301-7, 1991.
Article em Fr | MEDLINE | ID: mdl-1928848
ABSTRACT
The microhemagglutination-Treponema pallidum (MHA-TP) was automated with the Biomek 1000 workstation (Beckman). 3,000 serum specimens were studied with manual and automated procedures. Thirty nine serum specimens were found positive by both techniques, neither false positive nor false negative results were observed with the Biomek 1000 workstation. The statistical analyse showed no differences between the two procedures, variations were observed only with low titer sera. In conclusion, time saving and augmentation of productivity obtained by automation of this method could be interesting for a microbiology laboratory.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Testes de Hemaglutinação Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Fr Revista: Ann Biol Clin (Paris) Ano de publicação: 1991 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Testes de Hemaglutinação Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Fr Revista: Ann Biol Clin (Paris) Ano de publicação: 1991 Tipo de documento: Article