We retrospectively reviewed 31 hips with massive bone defect because of aseptic loosening of the acetabulum treated with impacted bone allografts with the Kerboull-type acetabular device. Mean age at surgery was 67.9 years, and mean duration of follow-up was 6.3 years. Thickness of the bone graft was evaluated by postoperative radiography. Seven hips showed radiological failure associated with 6 breakages of the device, and 24 hips showed stability on follow-up radiographs. Of the 13 hips showing a bone graft thickness of greater than 20 mm on follow-up radiographs, 7 were classified as failure group. If an acetabular reinforcement device with a bone graft of more than 20 mm thickness is necessary, then a structural allograft should be considered in the weight-bearing area of the support ring in addition to the morselized bone graft.