ABSTRACT
BACKGROUND AND

PURPOSE:

We are developing a technique for highly focused vocal cord irradiation in early glottic carcinoma to optimally treat a target volume confined to a single cord. This technique, in contrast with the conventional methods, aims at sparing the healthy vocal cord. As such a technique requires sub-mm daily targeting accuracy to be effective, we investigate the accuracy achievable with on-line kV-cone beam CT (CBCT) corrections. MATERIALS AND

METHODS:

CBCT scans were obtained in 10 early glottic cancer patients in each treatment fraction. The grey value registration available in X-ray volume imaging (XVI) software (Elekta, Synergy) was applied to a volume of interest encompassing the thyroid cartilage. After application of the thus derived corrections, residue displacements with respect to the planning CT scan were measured at clearly identifiable relevant landmarks. The intra- and inter-observer variations were also measured.

RESULTS:

While before correction the systematic displacements of the vocal cords were as large as 2.4+/-3.3mm (cranial-caudal population mean+/-SD Sigma), daily CBCT registration and correction reduced these values to less than 0.2+/-0.5mm in all directions. Random positioning errors (SD sigma) were reduced to less than 1mm. Correcting only for translations and not for rotations did not appreciably affect this accuracy. The residue random displacements partly stem from intra-observer variations (SD=0.2-0.6mm).

CONCLUSION:

The use of CBCT for daily image guidance in combination with standard mask fixation reduced systematic and random set-up errors of the vocal cords to <1mm prior to the delivery of each fraction dose. Thus, this facilitates the high targeting precision required for a single vocal cord irradiation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prega Vocal / Planejamento da Radioterapia Assistida por Computador / Sistemas On-Line / Neoplasias Laríngeas / Radioterapia de Intensidade Modulada / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Clinical_trials / Guideline Limite: Female / Humans / Male Idioma: En Revista: Radiother Oncol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Prega Vocal / Planejamento da Radioterapia Assistida por Computador / Sistemas On-Line / Neoplasias Laríngeas / Radioterapia de Intensidade Modulada / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Clinical_trials / Guideline Limite: Female / Humans / Male Idioma: En Revista: Radiother Oncol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Holanda