Thromboembolic complications during weaning from right ventricular assist device support.
Velik-Salchner, Corinna; Hoermann, Christoph; Hoefer, Daniel; Margreiter, Josef; Mair, Peter.
Afiliação
  • Velik-Salchner C; Department of Anaesthesiology and Intensive Care Medicine, Innsbruck Medical University, Anichstrasse 35, A-6020 Innsbruck, Austria. corinna.velik@i-med.ac.a
Anesth Analg ; 109(2): 354-7, 2009 Aug.
Article em En | MEDLINE | ID: mdl-19608801
ABSTRACT
We report on two patients who experienced life-threatening thromboembolic complications during prolonged weaning from Levitronix CentriMag right ventricular assist device support. Right ventricular assist device flow in both patients was reduced below 2 L/min for a period of 72 or 36 h to closely monitor patients' hemodynamics and echocardiography because the extent of right ventricular recovery was difficult to assess. Thrombus formation occurred despite adjusting the heparin dosage to achieve partial prothrombin time values between 60 and 70 s. Periods of reduced flow during Levitronix CentriMag support must be kept short, and an additional bolus dose of heparin should be considered.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tromboembolia / Coração Auxiliar Limite: Female / Humans / Middle aged Idioma: En Revista: Anesth Analg Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Áustria

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tromboembolia / Coração Auxiliar Limite: Female / Humans / Middle aged Idioma: En Revista: Anesth Analg Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Áustria