Influence of reduction in the elution times on HPLC glycohaemoglobin results.
Leon-Justel, Antonio; Santotoribio, Jose D; Dominguez-Pascual, Inmaculada; Delgado, Ana L; Macias, Carlos; Herrera, Maria T; Leal, Alfonso; Mangas, Miguel A; Acosta, Domingo; Guerrero, J M.
Afiliação
  • Leon-Justel A; Department of Clinical Biochemistry, Laboratory Buildings (4th Floor), Virgen del Rocío University Hospital, IBIS/CSIC/University of Seville, Seville, Spain. antonio.leonj.sspa@juntadeandalucia.es
Clin Biochem ; 42(15): 1582-4, 2009 Oct.
Article em En | MEDLINE | ID: mdl-19615990
ABSTRACT

OBJECTIVES:

To compare HPLC methods with short and long elution times for HbA1c measurement in blood.

METHODS:

Comparison between G7-Tosoh (1.2 min); Bio-Rad-Variant-II-Turbo (1.3 min) and Arkray-HA-8160 (2.9 min).

RESULTS:

Passing-Bablok regression equations were Y=0.17+0.96X; Y=-0.39+1.01X; Y=-0.40+1.0X and the means of the differences using Bland-Altman Plot were 0.02; -0.34; 0.32 for HA-8160/G7-Tosoh, HA-8160/Variant-II-Turbo and G7-Tosoh/Variant-II-Turbo, respectively.

CONCLUSIONS:

Faster elution methods had no problems on reproducibility of results obtained by slower elution methods.