ABSTRACT
The effects of radiation backscattered from the secondary collimators into the monitor chamber in an Elekta linac (producing 6 and 10 MV photon beams) are investigated using BEAMnrc Monte Carlo simulations. The degree and effects of this backscattered radiation are assessed by evaluating the changes to the calculated dose in the monitor chamber, and by determining a correction factor for those changes. Additionally, the fluence and energy characteristics of particles entering the monitor chamber from the downstream direction are evaluated by examining BEAMnrc phase-space data. It is shown that the proportion of particles backscattered into the monitor chamber is small (< 0.35%), for all field sizes studied. However, when the backscatter plate is removed from the model linac, these backscattered particles generate a noticeable increase in dose to the monitor chamber (up to approximately 2.4% for the 6 MV beam and up to 4.4% for the 10 MV beam). With its backscatter plate in place, the Elekta linac (operating at 6 and 10 MV) is subject to negligible variation of monitor chamber dose with field size. At these energies, output variations in photon beams produced by the clinical Elekta linear accelerator can be attributed to head scatter alone. Corrections for field-size-dependence of monitor chamber dose are not necessary when running Monte Carlo simulations of the Elekta linac operating at 6 and 10 MV.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radioterapia / Método de Monte Carlo Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Austrália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radioterapia / Método de Monte Carlo Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Australas Phys Eng Sci Med Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Austrália