Effects of radiofrequency probe application on irrigation fluid temperature in the wrist joint.
Sotereanos, Dean G; Darlis, Nickolaos A; Kokkalis, Zinon T; Zanaros, George; Altman, Gregory T; Miller, Mark Carl.
Afiliação
  • Sotereanos DG; Department of Orthopaedic Surgery, Hand and Upper Extremity Surgery, and the Orthopaedic Biomechanics Research Laboratory, Allegheny General Hospital, University of Pittsburgh, Pittsburgh, PA 15261, USA.
J Hand Surg Am ; 34(10): 1832-7, 2009 Dec.
Article em En | MEDLINE | ID: mdl-19910145
ABSTRACT

PURPOSE:

Radiofrequency (RF) probes used in wrist arthroscopy may raise joint fluid temperature, increasing the risk of capsular and ligamentous damage. The purposes of the current study were to measure joint fluid temperature during wrist arthroscopy with the use of RF probes, and to determine whether using an outlet portal will reduce the maximum temperature.

METHODS:

We performed wrist arthroscopy on 8 cadaveric arms. Ablation and coagulation cycles using RF probe were performed at documented locations within the joint. This was done for 60-second intervals on both the radial and ulnar side of the wrist, to mimic clinical practice. We used 4 fiberoptic phosphorescent probes to measure temperature (radial, ulnar, inflow-tube, and outflow-tube probes) and measured joint fluid temperature with and without outflow.

RESULTS:

There was a significant difference between wrists with and without outflow when examining maximum ablation temperatures (p < .002). All specimens showed higher maximum and average ablation temperatures without outflow. Maximum joint temperatures, greater than 60 degrees C, were observed in only no-outflow conditions.

CONCLUSIONS:

In performing RF ablation during wrist arthroscopy, the use of an outlet portal reduces the joint fluid temperature. Without an outlet portal, maximum temperatures can exceed desirable levels when using ablation; such temperatures have the potential to damage adjacent tissues. It is useful to maintain adequate outflow when using the radiofrequency probes during wrist arthroscopy.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroscopia / Líquido Sinovial / Articulação do Punho / Temperatura Corporal / Eletrocoagulação Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Hand Surg Am Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroscopia / Líquido Sinovial / Articulação do Punho / Temperatura Corporal / Eletrocoagulação Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: J Hand Surg Am Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Estados Unidos