Experimental verification of beam characteristics for cyclotron-based epithermal neutron source (C-BENS).
Appl Radiat Isot
; 69(12): 1642-5, 2011 Dec.
Article
em En
| MEDLINE
| ID: mdl-21463945
ABSTRACT
A cyclotron-based epithermal neutron source has been developed for boron neutron capture therapy. This system consists of a cyclotron accelerator producing 1.1-mA proton beams with an energy of 30 MeV, a beam transport system coupled with a beryllium neutron production target, and a beam-shaping assembly (BSA) with a neutron collimator. In our previous work, the BSA was optimized to obtain sufficient epithermal neutron fluxes of ~10(9) cm(-2) s(-1) using a Monte Carlo simulation code. In order to validate the simulation results, irradiation tests using multi-foil activation at the surface of a gamma-ray shield located behind the collimator and water phantom experiments using a collimated epithermal neutron beam were performed. It was confirmed experimentally that the intensity of the epithermal neutrons was 1.2×10(9) cm(-2) s(-1).
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Appl Radiat Isot
Assunto da revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Ano de publicação:
2011
Tipo de documento:
Article
País de afiliação: