Gas chromatography connected to multiple channel electrospray ionization mass spectrometry for the detection of volatile organic compounds.
Lee, C Y; Shiea, J.
Afiliação
  • Lee CY; Department of Chemistry, National Sun Yat-Sen University, Kaohsiung, 804 Taiwan.
Anal Chem ; 70(13): 2757-61, 1998 Jul 01.
Article em En | MEDLINE | ID: mdl-21644791
ABSTRACT
This work successfully connected gas chromatography (GC) to seven-channel electrospray ionization (ESI) mass spectrometry to separate and detect a mixture of volatile organic compounds. Gaseous analyte was eluted separately from a GC column and directed into the central channel of the ESI source. The analyte was protonated by ion-molecule reactions between the analyte and the ions which were generated by electrospraying the acidic solution through the outside six channels surrounding the central channel. Real-time analysis of the organic reaction involving volatile and thermally unstable compounds (dimethylhydrazine ⇌ azomethane + H(2)) was also achieved by continuously purging the air in the reaction vessel to the seven-channel ESI source.