Accuracy of volumetric measurement of teeth in vivo based on cone beam computer tomography.
Wang, Y; He, S; Yu, L; Li, J; Chen, S.
Afiliação
  • Wang Y; Department of Orthodontics, West China College of Stomatology, Sichuan University, Chengdu, China.
Orthod Craniofac Res ; 14(4): 206-12, 2011 Nov.
Article em En | MEDLINE | ID: mdl-22008300
ABSTRACT

OBJECTIVES:

to evaluate the accuracy of cone beam computed tomography (CBCT) for volumetric measurement of teeth, using micro-computed tomography (Micro-CT) as the reference standard. SETTING AND SAMPLE POPULATION The Department of Orthodontics at Sichuan University. The sample consisted of 27 maxillary and mandibular premolars of 15 patients, planned to be extracted for orthodontic treatment. MATERIAL AND

METHODS:

The 27 teeth were subjected to standardized CBCT scanning before extraction and Micro-CT scanning after extraction. From CBCT data, teeth were tissue segmented and then three-dimensionally (3D) reconstructed, while from Micro-CT data, teeth were 3D reconstructed directly. Tooth volumes were then calculated. The intra-observer repeatability and reproducibility of two observers and the overall between-instrument agreement of the measurements were evaluated using intra-class correlation coefficients (ICCs) and concordance correlation coefficient (CCC), respectively.

RESULTS:

The intra-observer repeatability was high for both observers. ICCs were 0.999 and 0.998, respectively. The reproducibility of the two observers was also high (ICC, 0.740). The overall between-instrument agreement of the measurements was good, and CCC was 0.993 and its lower 95% confidence interval was 0.989.

CONCLUSIONS:

The accuracy of the CBCT method for volumetric measurement of teeth in vivo is comparable to the Micro-CT method in vitro. The CBCT method has the potential possibility to be applied in studies on root resorption associated with orthodontic force. Further study is needed to prove the sensitivity of the method.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dente Pré-Molar / Tomografia Computadorizada de Feixe Cônico Limite: Adolescent / Adult / Child / Female / Humans / Male Idioma: En Revista: Orthod Craniofac Res Assunto da revista: ODONTOLOGIA / ORTODONTIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dente Pré-Molar / Tomografia Computadorizada de Feixe Cônico Limite: Adolescent / Adult / Child / Female / Humans / Male Idioma: En Revista: Orthod Craniofac Res Assunto da revista: ODONTOLOGIA / ORTODONTIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: China