The effect of sterilisation on the plasticity of multi-use Eschmann gum elastic bougies: a bench and manikin study.
Dawes, T J W; Ford, P N R.
Afiliação
  • Dawes TJ; Royal Devon and Exeter Hospital, Exeter, UK. timdawes@yahoo.co.uk
Anaesthesia ; 66(12): 1134-9, 2011 Dec.
Article em En | MEDLINE | ID: mdl-22074029
ABSTRACT
We investigated the deterioration in plasticity of the multi-use Portex Tracheal Guide ('bougie') with repeated sterilisation. Six bougies were prepared by washing them between 0 and 100 times (in accordance with the manufacturer's guidelines). Two tests were employed a bench test in which rapid serial photographs were taken of the bougies uncoiling from a preformed curve; and a manikin-based test in which the six bougies were used in a simulated difficult airway. The bench test demonstrated a progressive deterioration in plasticity with repeated washing. However, the manikin-based test showed no significant difference between bougies in the incidence of oesophageal placement (p = 0.74). Time to placement differed significantly only between the two most-washed bougies but was broadly similar. We suggest therefore that the manufacturer's limit of five washings may be unnecessarily cautious.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esterilização / Intubação Intratraqueal / Manequins Limite: Humans Idioma: En Revista: Anaesthesia Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esterilização / Intubação Intratraqueal / Manequins Limite: Humans Idioma: En Revista: Anaesthesia Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Reino Unido