MGuard mesh-covered stent for treatment of ST-segment elevation myocardial infarction with high thrombus burden despite manual aspiration.
Romaguera, Rafael; Gómez-Hospital, Joan A; Sánchez-Elvira, Guillermo; Gómez-Lara, Josep; Ferreiro, José L; Roura, Gerard; Gracida, Montserrat; Homs, Silvia; Teruel, Luis; Cequier, Angel.
Afiliação
  • Romaguera R; Heart Diseases Institute, Bellvitge University Hospital-IDIBELL, University of Barcelona, Barcelona, Spain. rafaromaguera@gmail.com
J Interv Cardiol ; 26(1): 1-7, 2013 Feb.
Article em En | MEDLINE | ID: mdl-23419104
ABSTRACT

OBJECTIVES:

To assess the usefulness of the MGuard stent in patients with ST-segment elevation myocardial infarction (STEMI) in whom a high thrombus burden persists after manual aspiration.

BACKGROUND:

In some patients with STEMI, a high thrombus burden may persist after manual aspiration. These patients may be at high risk of distal embolization and therefore impaired myocardial reperfusion. The MGuard is a novel mesh-covered stent designed to minimize thrombus embolization.

METHODS:

Single-arm, prospective registry of patients with STEMI and high thrombus burden after aggressive thrombus aspiration treated with the MGuard stent. High thrombus burden was defined as thrombus burden grade 4 or 5 according to the TIMI score. Lesions with a side branch ≥2 mm and patients with cardiogenic shock were not included. The study end-points were proportion of final TIMI 3 flow, normal myocardial blush, and complete ST-segment resolution.

RESULTS:

Fifty-six patients were included. After MGuard stent implantation >85% of cases had thrombus score = 0. Final TIMI 3 flow was achieved in 82% of cases, normal myocardial blush in 55%, and complete ST-segment resolution in 59%. Occlusion of a side branch (<2 mm) occurred in 2 cases (3.5%), embolization to a distal branch in 5 cases (8.9%), and transient no-reflow in 4 cases (7.1%). Major adverse cardiac events rate at 9 months was 3.6%, including 1 definite acute stent thrombosis and 1 target-vessel revascularization.

CONCLUSIONS:

The MGuard stent may be useful to prevent distal embolization in patients with STEMI and high thrombus burden despite mechanical aspiration.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Stents / Trombectomia / Dispositivos de Proteção Embólica / Infarto do Miocárdio Tipo de estudo: Guideline / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Stents / Trombectomia / Dispositivos de Proteção Embólica / Infarto do Miocárdio Tipo de estudo: Guideline / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Espanha