Detailed analysis of polymer response to delivery balloon expansion of drug-eluting stents versus bare metal stents.
Denardo, Scott J; Carpinone, Paul L; Vock, David M; Tcheng, James E; Phillips, Harry R; Willenberg, Bradley J; Batich, Christopher D; Pepine, Carl J.
Afiliação
  • Denardo SJ; Division of Cardiovascular Medicine, Duke University Medical Center, Durham, NC 27710, USA. scott.denardo@duke.edu
EuroIntervention ; 9(3): 389-97, 2013 Jul.
Article em En | MEDLINE | ID: mdl-23872653
ABSTRACT

AIMS:

We sought to describe the response of the polymer surface of drug-eluting stents (DES) to delivery balloon expansion, including quantitation of any resulting detached microparticles. METHODS AND

RESULTS:

We expanded the US Food and Drug Administration (FDA)-approved first- and second-generation DES in a vacuum filtration system and used optical and scanning electron microscopy to image the polymer surface, filters and delivery balloons. DES were expanded under a range of conditions, from in vitro conditions used for FDA regulatory submissions to human in vivo conditions. Dispersive Raman spectroscopy was used for definitive identification of microparticles. All polymer surfaces were topographically disturbed over an average of 4.6%-100% of the surface area imaged. Disturbances ranged from deformation (including peeling) to complete delamination. The dimensions of detached microparticles were 2-350 µm. The extent and nature of surface disturbances and microparticles were primarily a function of polymer composition (p<0.001 for 8/10 disturbance types/locations) and were independent of expansion condition (p=0.100 to 0.989 for 9/10 disturbance types/locations).

CONCLUSIONS:

Balloon expansion of first- and second-generation DES disturbs the polymer surface and can cause detachment of microparticles; each is functionally related to the specific polymer but not to expansion condition. Disturbance "roughness" and detached microparticles may contribute to DES limitations.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polímeros / Angioplastia Coronária com Balão / Stents / Stents Farmacológicos / Metais Idioma: En Revista: EuroIntervention Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polímeros / Angioplastia Coronária com Balão / Stents / Stents Farmacológicos / Metais Idioma: En Revista: EuroIntervention Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos