Status and characterization of the lithium beam diagnostic on DIII-D.
Stoschus, H; Thomas, D M; Hudson, B; Watkins, M; Finkenthal, D F; Moyer, R A; Osborne, T H.
Afiliação
  • Stoschus H; Oak Ridge Institute for Science and Education, Oak Ridge, Tennessee 37831-0117, USA.
Rev Sci Instrum ; 84(8): 083503, 2013 Aug.
Article em En | MEDLINE | ID: mdl-24007061
ABSTRACT
The 30 keV lithium beam diagnostic on DIII-D is suitable to measure both the radial electron density and poloidal magnetic field profiles in the pedestal. The refurbished system features a new setup to measure the Doppler shift allowing accurate alignment of the spectral filters. The injector has been optimized to generate a stable lithium neutral beam with a current of I = 15-20 mA and a diameter of 1.9 ± 0.1 cm measured by beam imaging. The typical temporal resolution is Δt = 1-10 ms and the radial resolution of ΔR = 5 mm is given by the optical setup. A new analysis technique based on fast Fourier transform avoids systematic error contributions from the digital lock-in analysis and accounts intrinsically for background light correction. Latest upgrades and a detailed characterization of the system are presented. Proof-of-principle measurements of the poloidal magnetic field with a statistical error of typically 2% show a fair agreement with the predictions modeled with the Grad-Shafranov equilibrium solver EFIT within 4%.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos