Ability of the third-generation FloTrac/Vigileo software to track changes in cardiac output in cardiac surgery patients: a polar plot approach.
Desebbe, Olivier; Henaine, Roland; Keller, Geoffray; Koffel, Catherine; Garcia, Hans; Rosamel, Pascal; Obadia, Jean-Francois; Bastien, Olivier; Lehot, Jean-Jacques; Haftek, Marek; Critchley, Lester A H.
Afiliação
  • Desebbe O; Department of Anesthesiology and Intensive Care, Louis Pradel Hospital; Laboratory EA4169, Claude Bernard Lyon 1 University, Lyon, France. Electronic address: oldesebbe@yahoo.com.
J Cardiothorac Vasc Anesth ; 27(6): 1122-7, 2013 Dec.
Article em En | MEDLINE | ID: mdl-24011911
OBJECTIVE: To evaluate the ability of the third-generation (3.01) of FloTrac/Vigileo monitor (Edwards Lifesciences, Irvine, CA) to follow variations in cardiac output (∆CO) using the new polar plot approach. DESIGN: Prospective interventional study. SETTING: Single hospital university study. PARTICIPANTS: Twenty-five patients referred for cardiac surgery. INTERVENTIONS: CO was measured simultaneously by 3 to 5 bolus thermodilution (COtd measurements), using a pulmonary artery catheter and by arterial pulse contour analysis, using the FloTrac/Vigileo (COvi). Data were collected at eight time points: before incision, after sternotomy, before and after protamine sulfate infusion, at the start of sternal closure, at the end of surgery, on arrival to intensive care unit, and after a standardized volume expansion with 500 mL of hetastarch 6%. MEASUREMENTS AND MAIN RESULTS: One-hundred thirty-five pairs of CO data were collected; the mean bias of all CO measurements corrected for repeated measures was 0.2 L/min with limits of agreements of -3.3 L/min and +2.9 L/min. The percentage error was 66.5%. The polar plot analysis included 71 significant ∆CO and showed a mean polar angle of -3.4 degrees with 95% polar percentage error equivalent limits of -61 to 55; 69% of analysed data points fell within the 30-degree limits and provided a correct polar concordance rate. CONCLUSIONS: Third-generation FloTrac/Vigileo software still lacks the accuracy to reliably detect changes in cardiac output (∆CO) in cardiac surgery. Improvements to FloTrac/Vigileo CO algorithm and software still are needed in this particular setting.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Débito Cardíaco / Monitorização Intraoperatória / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cardiothorac Vasc Anesth Assunto da revista: ANESTESIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Software / Débito Cardíaco / Monitorização Intraoperatória / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cardiothorac Vasc Anesth Assunto da revista: ANESTESIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article