Reproducibility study of four-dimensional flow MRI of arterial and portal venous liver hemodynamics: influence of spatio-temporal resolution.
Stankovic, Zoran; Jung, Bernd; Collins, Jeremy; Russe, Maximilian F; Carr, James; Euringer, Wulf; Stehlin, Lena; Csatari, Zoltan; Strohm, Peter C; Langer, Mathias; Markl, Michael.
Afiliação
  • Stankovic Z; Department of Radiology, Northwestern University, Chicago, Illinois, USA; Department of Biomedical Engineering, Northwestern University, Chicago, Illinois, USA; Department of Diagnostic Radiology and Medical Physics, University Medical Center Freiburg, Freiburg, Germany.
Magn Reson Med ; 72(2): 477-84, 2014 Aug.
Article em En | MEDLINE | ID: mdl-24018798
ABSTRACT

PURPOSE:

To evaluate influence of variation in spatio-temporal resolution and scan-rescan reproducibility on three-dimensional (3D) visualization and quantification of arterial and portal venous (PV) liver hemodynamics at four-dimensional (4D) flow MRI.

METHODS:

Scan-rescan reproducibility of 3D hemodynamic analysis of the liver was evaluated in 10 healthy volunteers using 4D flow MRI at 3T with three different spatio-temporal resolutions (2.4 × 2.0 × 2.4 mm(3), 61.2 ms; 2.5 × 2.0 × 2.4 mm(3), 81.6 ms; 2.6 × 2.5 × 2.6 mm(3), 80 ms) and thus different total scan times. Qualitative flow analysis used 3D streamlines and time-resolved particle traces. Quantitative evaluation was based on maximum and mean velocities, flow volume, and vessel lumen area in the hepatic arterial and PV systems.

RESULTS:

4D flow MRI showed good interobserver variability for assessment of arterial and PV liver hemodynamics. 3D flow visualization revealed limitations for the left intrahepatic PV branch. Lower spatio-temporal resolution resulted in underestimation of arterial velocities (mean 15%, P < 0.05). For the PV system, hemodynamic analyses showed significant differences in the velocities for intrahepatic portal vein vessels (P < 0.05). Scan-rescan reproducibility was good except for flow volumes in the arterial system.

CONCLUSION:

4D flow MRI for assessment of liver hemodynamics can be performed with low interobserver variability and good reproducibility. Higher spatio-temporal resolution is necessary for complete assessment of the hepatic blood flow required for clinical applications.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Velocidade do Fluxo Sanguíneo / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Técnicas de Imagem de Sincronização Cardíaca / Artéria Hepática / Veias Hepáticas / Circulação Hepática Tipo de estudo: Diagnostic_studies / Qualitative_research Limite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Velocidade do Fluxo Sanguíneo / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Técnicas de Imagem de Sincronização Cardíaca / Artéria Hepática / Veias Hepáticas / Circulação Hepática Tipo de estudo: Diagnostic_studies / Qualitative_research Limite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Alemanha