Your browser doesn't support javascript.
loading
First-in-human evaluation of a bioabsorbable polymer-coated sirolimus-eluting stent: imaging and clinical results of the DESSOLVE I Trial (DES with sirolimus and a bioabsorbable polymer for the treatment of patients with de novo lesion in the native coronary arteries).
Ormiston, John; Webster, Mark; Stewart, James; Vrolix, Mathias; Whitbourn, Robert; Donohoe, Dennis; Knape, Charlene; Lansky, Alexandra; Attizzani, Guilherme F; Fitzgerald, Peter; Kandzari, David E; Wijns, William.
Afiliação
  • Ormiston J; Mercy Angiography Unit, Auckland, New Zealand; Auckland City Hospital, Auckland, New Zealand. Electronic address: johno@mercyangiography.co.nz.
JACC Cardiovasc Interv ; 6(10): 1026-34, 2013 Oct.
Article em En | MEDLINE | ID: mdl-24055443
ABSTRACT

OBJECTIVES:

This first-in-human multicenter study sought to examine prospectively the safety and efficacy of a new, cobalt chromium thin-strut, coronary absorbable polymer-coated, sirolimus-eluting stent.

BACKGROUND:

Bioabsorbable polymers on drug-eluting stents may lower the long-term risks of inflammation, delayed healing, and adverse events.

METHODS:

We enrolled patients with symptomatic coronary artery disease with stable or unstable angina pectoris and >50% diameter stenosis, amenable to coverage with a ≤23-mm long stent in a vessel 2.5 to 3.5 mm in diameter. All patients received dual antiplatelet therapy after implantation. Patients, in groups of 10, underwent repeat angiography, intravascular ultrasound, and optical coherence tomography at 4, 6, or 8 months, and all patients were seen or contacted at 18 months of follow-up.

RESULTS:

The median (range) in-stent late lumen loss (LLL) was 0.03 mm (-0.22 to 0.21 mm), 0.10 mm (-0.03 to 1.2 mm), and 0.08 mm (-0.01 to 0.28 mm), at 4, 6, and 8 months, respectively. At 18 months, the median in-stent LLL was 0.08 mm (-0.30 to 0.46 mm). On optical coherence tomography, the proportion of uncovered stent struts decreased from a median of 7.3% (range 0.4% to 46.3%) at 4 months to 0% (range 0% to 3.4%) at 18 months. The percentage of neointimal volume obstruction by intravascular ultrasound increased from a median of 5.3% to 9.1% between 4 and 6 months and remained nearly unchanged thereafter through 18 months of follow-up. The only recorded major adverse cardiac event was a myocardial infarction.

CONCLUSIONS:

At 18 months of follow-up, this absorbable polymer-coated, cobalt chromium sirolimus-eluting stent was associated with a low and stable in-stent LLL, complete strut coverage, and no stent thrombosis. (First-In-Human Trial of the MiStent Drug-Eluting Stent [DES] in Coronary Artery Disease [DESSOLVE-I]; NCT01247428).
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polímeros / Doença da Artéria Coronariana / Fármacos Cardiovasculares / Sirolimo / Implantes Absorvíveis / Estenose Coronária / Stents Farmacológicos / Intervenção Coronária Percutânea Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged80 País/Região como assunto: Europa / Oceania Idioma: En Revista: JACC Cardiovasc Interv Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Polímeros / Doença da Artéria Coronariana / Fármacos Cardiovasculares / Sirolimo / Implantes Absorvíveis / Estenose Coronária / Stents Farmacológicos / Intervenção Coronária Percutânea Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged80 País/Região como assunto: Europa / Oceania Idioma: En Revista: JACC Cardiovasc Interv Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article