Very low field magnetic resonance imaging with spintronic sensors.
Herreros, Q; Dyvorne, H; Campiglio, P; Jasmin-Lebras, G; Demonti, A; Pannetier-Lecoeur, M; Fermon, C.
Afiliação
  • Herreros Q; DSM∕IRAMIS∕SPEC-CNRS URA 2464, CEA Saclay, F-91191 Gif sur Yvette Cedex, France.
Rev Sci Instrum ; 84(9): 095116, 2013 Sep.
Article em En | MEDLINE | ID: mdl-24089875
A very low field magnetic resonance imaging (MRI) setup based on magnetoresistive-superconducting mixed sensors is presented. A flux transformer is used to achieve coupling between the sample to image and the mixed sensor. The novel detector was implemented in a spin echo MRI experiment, exposing the mixed sensor to RF pulses without use of any RF switch. The performance of the novel detector is given in terms of signal-to-noise ratio and is compared with classical tuned coils.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Razão Sinal-Ruído Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2013 Tipo de documento: Article País de afiliação: França