[Study on the effect of target thickness on output spectrum in transmission anode micro X-ray tube by MC method].
Guang Pu Xue Yu Guang Pu Fen Xi
; 33(8): 2231-4, 2013 Aug.
Article
em Zh
| MEDLINE
| ID: mdl-24159883
ABSTRACT
The transmission anode micro X-ray tube is a key component of energy dispersive X-ray Fluorescence Analysis (EDXRFA) as exciting source. So in EDXRFA we hope to get the primary X-ray spectrum which is simply distributed. In the present paper, the authors used the Monte Carlo (MC) method to study the spectrum character of the transmission anode micro X-ray tube with different thickness of target. We cut the spectrum into high energy (5keV-50keV) part and low energy (< 5 keV) part and compared them to each other. The result showed that the flux of high energy part is big and the flux of low energy part is small when the electron energy is 50 keV and the target is 4 microm thick Ag. This is better for EDXRFA.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
Zh
Revista:
Guang Pu Xue Yu Guang Pu Fen Xi
Ano de publicação:
2013
Tipo de documento:
Article
País de afiliação:
China