Your browser doesn't support javascript.
loading
Dedicated bifurcation paclitaxel-eluting stent BiOSS Expert® in the treatment of distal left main stem stenosis.
Bil, Jacek; Gil, Robert J; Vassilev, Dobrin; Rzezak, Jaroslaw; Kulawik, Tomasz; Pawlowski, Tomasz.
Afiliação
  • Bil J; Department of Invasive Cardiology, Central Clinical Hospital of the Ministry of Interior, Warsaw, Poland.
J Interv Cardiol ; 27(3): 242-51, 2014 Jun.
Article em En | MEDLINE | ID: mdl-24708143
ABSTRACT

OBJECTIVES:

The aim of this study was to assess prospectively the effectiveness and safety profile of distal left main stem (LMS) stenosis treatment with dedicated bifurcation paclitaxel-eluting stent BiOSS Expert®.

BACKGROUND:

Angioplasty of distal LMS stenosis is always a high-risk procedure, and optimal treatment is uncertain.

METHODS:

This was a prospective international 2-center study, which enrolled patients with non-ST-elevation acute coronary syndrome (NSTE-ACS) or stable angina with distal left main stenosis. All patients were treated with the dedicated bifurcation stent BiOSS Expert®. Provisional T-stenting was the obligatory strategy. Angiographic control was performed after 12 months. The primary end-point was cumulative rate of death, myocardial infarction (MI), and target lesion revascularization (TLR) at 12 months. Angiographic end-points included late lumen loss, percent diameter stenosis, and binary restenosis rate.

RESULTS:

A total of 54 patients with distal LMS stenosis were enrolled. Seven patients (13%) were enrolled during NSTE-ACS, 77.8% were hypertensive, 27.8% were diabetic, 51.9% had previous MI, 53.7% underwent prior percutaneous coronary intervention, and 16.7% coronary artery bypass graft. The mean SYNTAX score was 21.52 ± 6.50. The device success rate was 100%. The mean BiOSS Expert stent parameters were as follows 4.07 ± 0.26 mm × 3.36 ± 0.26 mm × 16.61 ± 1.72 mm and in side branch the other stent (classical drug-eluting stent) was implanted in 25.9% of cases. The overall TLR was 9.3%. There were no death, stent thrombosis, or acute MI. In the univariate regression analysis, the only factor associated with higher risk for TLR was the SYNTAX score value.

CONCLUSIONS:

The dedicated bifurcation stent BiOSS Expert® proved to be a feasible device, with promising safety and long-term clinical effectiveness in the treatment of distal LMS stenosis.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Paclitaxel / Estenose Coronária / Stents Farmacológicos / Angina Estável Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Polônia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angioplastia Coronária com Balão / Paclitaxel / Estenose Coronária / Stents Farmacológicos / Angina Estável Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Polônia