Evaluation of a minimally invasive procedure for sacroiliac joint fusion - an in vitro biomechanical analysis of initial and cycled properties.
Lindsey, Derek P; Perez-Orribo, Luis; Rodriguez-Martinez, Nestor; Reyes, Phillip M; Newcomb, Anna; Cable, Alexandria; Hickam, Grace; Yerby, Scott A; Crawford, Neil R.
Afiliação
  • Lindsey DP; SI-BONE, Inc., San Jose, CA, USA.
  • Perez-Orribo L; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Rodriguez-Martinez N; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Reyes PM; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Newcomb A; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Cable A; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Hickam G; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
  • Yerby SA; SI-BONE, Inc., San Jose, CA, USA.
  • Crawford NR; Spinal Biomechanics Research Laboratory, Barrow Neurological Institute, St Joseph's Hospital and Medical Center, Phoenix, AZ, USA.
Med Devices (Auckl) ; 7: 131-7, 2014.
Article em En | MEDLINE | ID: mdl-24868175
ABSTRACT

INTRODUCTION:

Sacroiliac (SI) joint pain has become a recognized factor in low back pain. The purpose of this study was to investigate the effect of a minimally invasive surgical SI joint fusion procedure on the in vitro biomechanics of the SI joint before and after cyclic loading.

METHODS:

SEVEN CADAVERIC SPECIMENS WERE TESTED UNDER THE FOLLOWING CONDITIONS intact, posterior ligaments (PL) and pubic symphysis (PS) cut, treated (three implants placed), and after 5,000 cycles of flexion-extension. The range of motion (ROM) in flexion-extension, lateral bending, and axial rotation was determined with an applied 7.5 N · m moment using an optoelectronic system. Results for each ROM were compared using a repeated measures analysis of variance (ANOVA) with a Holm-Sidák post-hoc test.

RESULTS:

Placement of three fusion devices decreased the flexion-extension ROM. Lateral bending and axial rotation were not significantly altered. All PL/PS cut and post-cyclic ROMs were larger than in the intact condition. The 5,000 cycles of flexion-extension did not lead to a significant increase in any ROMs.

DISCUSSION:

In the current model, placement of three 7.0 mm iFuse Implants significantly decreased the flexion-extension ROM. Joint ROM was not increased by 5,000 flexion-extension cycles.