Your browser doesn't support javascript.
loading
Results of the U.S. Food and Drug Administration continued access clinical trial of the GORE HELEX septal occluder for secundum atrial septal defect.
Javois, Alexander J; Rome, Jonathan J; Jones, Thomas K; Zahn, Evan M; Fleishman, Craig E; Pignatelli, Ricardo H; Latson, Larry A.
Afiliação
  • Javois AJ; Advocate Children's Hospital, Oak Lawn, Illinois; University of Illinois Medical College, Chicago, Illinois. Electronic address: alexander.javois@advocatehealth.com.
  • Rome JJ; Children's Hospital of Philadelphia, Philadelphia, Pennsylvania.
  • Jones TK; Seattle Children's Hospital, Seattle, Washington.
  • Zahn EM; Heart Institute and Department of Pediatrics at Cedars-Sinai Medical Center, Los Angeles, California.
  • Fleishman CE; Arnold Palmer Hospital for Children, Orlando, Florida.
  • Pignatelli RH; Texas Children's Hospital, Houston, Texas.
  • Latson LA; Joe DiMaggio Children's Hospital, Hollywood, Florida.
JACC Cardiovasc Interv ; 7(8): 905-12, 2014 Aug.
Article em En | MEDLINE | ID: mdl-25147036
ABSTRACT

OBJECTIVES:

This report describes the immediate, 1-, and 5-year follow-up results of the U.S. Food and Drug Administration Continued Access clinical trial of the GORE HELEX Septal Occluder (W. L. Gore & Associates, Inc., Flagstaff, Arizona) for closure of secundum atrial septal defect.

BACKGROUND:

The trial was conducted between May 2003 and August 2006 to allow continued enrollment in a trial of the investigational device during review of data from the Pivotal Trial. Devices with hydrophilic coating on the expanded polytetrafluoroethylene to improve echocardiographic visualization were first used in this trial.

METHODS:

A total of 137 devices were implanted in 137 patients at 13 US institutions. Evaluations were scheduled at 1, 6, and 12 months for the initial trial and at 36 and 60 months for a later extension of the trial in those who consented to longer term evaluations.

RESULTS:

Twelve-month follow-up was completed on 122 of 126 patients with implantations, and 5-year follow-up on 83 of 95 patients who agreed to the trial extension. The overall clinical success rate was 96.7%, and the major adverse event rate 3.6%. Wire frame fractures were seen in 11.7% of patients with no clinical symptoms. A trivial, clinically insignificant leak was seen, or could not be ruled out, in 26.6% of patients at the 5-year evaluation, but no clinically significant leaks were seen. No patient experienced an erosion or sudden catastrophic event.

CONCLUSIONS:

The immediate, 1-, and 5-year follow-up outcomes of the Continued Access clinical trial continue to demonstrate that the GORE HELEX Septal Occluder is a safe and effective transcatheter occluder for repair of ostium secundum atrial septal defect.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: United States Food and Drug Administration / Cateterismo Cardíaco / Materiais Revestidos Biocompatíveis / Dispositivo para Oclusão Septal / Comunicação Interatrial Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged País/Região como assunto: America do norte Idioma: En Revista: JACC Cardiovasc Interv Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: United States Food and Drug Administration / Cateterismo Cardíaco / Materiais Revestidos Biocompatíveis / Dispositivo para Oclusão Septal / Comunicação Interatrial Tipo de estudo: Clinical_trials / Diagnostic_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged País/Região como assunto: America do norte Idioma: En Revista: JACC Cardiovasc Interv Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article