ABSTRACT

BACKGROUND:

TAVR has become an established treatment for severe symptomatic aortic stenosis in patients with high surgical risk. The latest generation of the balloon-expandable Edwards Sapien device, the Sapien 3, together with its new transfemoral Commander delivery system has been designed to reduce paravalvular regurgitation and vascular access site complications.

OBJECTIVES:

To evaluate procedural results and short term outcome with the third generation Sapien 3 device.

METHODS:

We retrospectively evaluated 125 consecutive TAVR patients and analyzed the first 51 patients in whom we implanted the new Sapien 3 device via transfemoral access.

RESULTS:

In patients implanted with the Sapien 3 device significant residual paravalvular regurgitation after TAVR was virtually absent with the vast majority having none or trace postinterventional aortic regurgitation on angiography or echocardiography (92.2% and 80.4% respectively). None of the patients had more than mild paravalvular regurgitation. Major vascular access site complications or major bleeding according to the VARC II criteria were not observed in our cohort, minor vascular complications and minor bleeding occurred in 7.8% and 5.9% respectively. If vascular complications occurred, they were related to closure device failure. Thirty day outcome showed a 1.9% major stroke rate and 3.9% death rate. However, we observed a 25.5% permanent pacemaker rate in our Sapien 3 cohort.

CONCLUSIONS:

Implantation of the new third generation Sapien 3 device resulted in excellent procedural and short term outcome. Significant paravalvular regurgitation was virtually absent. However, the increased rate of postinterventional pacemaker implantations needs to be analyzed in a larger cohort of patients.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Bioprótese / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged80 / Female / Humans / Male Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Bioprótese / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Aged80 / Female / Humans / Male Idioma: En Revista: J Interv Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Alemanha